Color Doppler regurgitant characteristics of normal mechanical mitral valve prostheses in vitro.

Author:

Baumgartner H1,Khan S1,DeRobertis M1,Czer L1,Maurer G1

Affiliation:

1. Division of Cardiology and Cardiovascular Surgery, Cedars-Sinai Medical Center, Los Angeles, Calif.

Abstract

BACKGROUND To evaluate normal regurgitant characteristics of St. Jude (SJ) and Medtronic-Hall (MH) mitral valves, four sizes (25-31 mm) of each were studied in a pulsatile flow model. METHODS AND RESULTS Regurgitant flow was measured by flowmeter at left ventricular pressures of 80, 130, and 180 mm Hg. Peak regurgitant flow rates ranged from 6.2 to 12.7 cm3/sec in SJ valves and from 7.9 to 17.5 cm3/sec in MH valves. Regurgitant orifice areas calculated from the Doppler continuity equation ranged from 1.6 to 2.0 mm2 in SJ valves and from 2.2 to 2.9 mm2 in MH valves. Regurgitant volumes across the closed valve at a left ventricular pressure of 130 mm Hg were normalized to an ejection time of 280 msec and ranged from 1.5 to 1.9 cm3 in SJ valves and from 2.1 to 2.8 cm3 in MH valves. Jets were imaged by color Doppler in six rotational planes, and jet size and morphology were compared with those of regurgitant jets from circular orifices with sizes comparable to the calculated prosthetic valve regurgitant orifices (1.1-3.1 mm2). SJ valves showed two converging jets from the pivot points, one central jet, and a variable number of peripheral jets. The mean color jet area derived from the six image planes ranged from 1.6 to 5.3 cm2. Aliasing occurred only close to the valve (maximal distance 0.5-2.0 cm). MH valves showed a large central jet with a maximal length of aliased flow between 2.0 and 5.5 cm. Depending on valve size, driving pressure, and image plane, one or two small peripheral jets were found. These jets did not show aliasing in any case. The mean color jet area ranged from 5.1 to 11.0 cm2. Jets originating from circular orifices of comparable size showed jet areas from 5.5 to 13.9 cm2 and aliasing distances from 3.3 to 7.3 cm. At similar regurgitant orifice areas, driving pressures, and regurgitant flows, the measured color areas and aliasing distances were smallest in SJ valves, larger in MH valves, and largest in simple circular orifices. CONCLUSIONS Large, complex regurgitant jets can be found in normal closed SJ and MH valves by color Doppler, although regurgitant flow volume is minimal. Jet size and velocity distribution differs markedly between SJ valves, MH valves, and circular orifices, even with comparable driving pressure, regurgitant orifice area, and regurgitant volume. The characteristic patterns of normal regurgitation must be recognized to avoid incorrect diagnoses of pathological regurgitation in SJ and MH prosthetic valves. MH valves should not be removed solely on the basis of a central regurgitant jet with a long aliasing distance. Peripheral jets in MH valves and all jets in SJ valves should be considered normal as long as no or only minimal aliasing is present. In contrast, peripheral jets with significant aliasing may represent strong evidence of pathological regurgitation.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 16 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Prosthetic heart valves: Types and echocardiographic evaluation;International Journal of Cardiology;2007-11

2. Herzklappenprothesen;Echokardiographie;2007

3. Prosthetic Valves;Atlas of Intraoperative Transesophageal Echocardiography;2007

4. Hemolysis in Mechanical Bileaflet Prostheses: Experience With the Bicarbon Valve;The Annals of Thoracic Surgery;2006-04

5. Tips und Tricks bei Aortenklappenvitien und Aortenklappenersatz;Angewandte Echokardiographie;2000

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3