Abstract

Background— Precise knowledge of left atrial appendage (LAA) orifice size is crucial for correct sizing of LAA closure devices. The aim of the present study was to determine the performance of real-time 3D transesophageal echocardiography (RT3DTEE) for LAA orifice size assessment, compared with 2D transesophageal echocardiography (2DTEE), and to investigate the impact of atrial fibrillation (AF) on LAA orifice size. Methods and Results— One hundred thirty-seven patients (38 control subjects, 31 with paroxysmal AF, 38 with persistent AF and 30 with permanent AF) underwent 2DTEE and RT3DTEE. Both techniques were used to measure LAA orifice area. Clinically-indicated 64-slice computed tomography (CT) was used as reference technique in 46 patients. Two-dimensional TEE underestimated LAA orifice area, compared with RT3DTEE (1.99±0.94 cm 2 versus 3.05±1.27 cm 2 ; P <0.001). RT3DTEE showed higher correlation with CT for the assessment of LAA orifice area, compared with 2DTEE ( r =0.92; 95% confidence interval, 0.85 to 0.95, versus r =0.72; 95% confidence interval, 0.54 to 0.83, respectively). At Bland–Altman analysis, RT3DTEE and 2DTEE underestimated LAA orifice area, compared with CT. However, RT3DTEE showed smaller bias (0.07 cm 2 versus 0.72 cm 2 ) and narrower limits of agreement (−0.71 to 0.85 cm 2 versus −0.58 to 2.02 cm 2 ) with CT, compared with 2DTEE. Among AF patients, a progressive increase in RT3DTEE-derived LAA orifice area was observed with increasing frequency of AF ( P <0.001). At multivariate analysis, AF and left atrial volume index ( P <0.001 for both) were independently associated with RT3DTEE-derived LAA orifice area. Conclusions— RT3DTEE is more accurate than 2DTEE for the assessment of LAA orifice size. A progressive increase in LAA orifice area is observed with increasing frequency of AF.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Cardiology and Cardiovascular Medicine,Radiology Nuclear Medicine and imaging

Cited by 175 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3