Affiliation:
1. From the Department of Cardiology, Athens Euroclinic, Athens, Greece (D.G.K., I.P.); Fluid Research, Athens, Greece (A.T.); Medical and Radiation Physics, Department of Radiology, University of Athens, Greece (I.P., E.P.E.); and the School of Engineering and Mathematical Sciences, The City University, London, United Kingdom (M.G., N.K.).
Abstract
Background—
The ideal bifurcation stenting technique is not established, and data on the hemodynamic characteristics at stented bifurcations are limited.
Methods and Results—
We used computational fluid dynamics analysis to assess hemodynamic parameters known affect the risk of restenosis and thrombosis at coronary bifurcations after the use of various single- and double-stenting techniques. We assessed the distributions and surface integrals of the time averaged wall shear stress (TAWSS), oscillatory shear index (OSI), and relative residence time (t
r
). Single main branch stenting without side branch balloon angioplasty or stenting provided the most favorable hemodynamic results (integrated values of TAWSS=4.13·10
−4
N, OSI=7.52·10
−6
m
2
, t
r
=5.57·10
−4
m
2
/Pa) with bifurcational area subjected to OSI values >0.25, >0.35, and >0.45 calculated as 0.36 mm
2
, 0.04 mm
2
, and 0 mm
2
, respectively. Extended bifurcation areas subjected to these OSI values were seen after T-stenting: 0.61 mm
2
, 0.18 mm
2
, and 0.02 mm
2
, respectively. Among the considered double-stenting techniques, crush stenting (integrated values of TAWSS=1.18·10
−4
N, OSI=7.75·10
−6
m
2
, t
r
=6.16·10
−4
m
2
/Pa) gave the most favorable results compared with T-stenting (TAWSS=0.78·10
−4
N, OSI=10.40·10
−6
m
2
, t
r
=6.87·10
−4
m
2
/Pa) or the culotte technique (TAWSS=1.30· 10
−4
N, OSI=9.87·10
−6
m
2
, t
r
=8.78·10
−4
m
2
/Pa).
Conclusions—
In the studied models of computer simulations, stenting of the main branch with our without balloon angioplasty of the side branch offers hemodynamic advantages over double stenting. When double stenting is considered, the crush technique with the use of a thin-strut stent may result in improved immediate hemodynamics compared with culotte or T-stenting.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Cardiology and Cardiovascular Medicine
Reference37 articles.
1. Simple or complex stenting for bifurcation coronary lesions: a patient-level pooled-analysis of the Nordic Bifurcation Study and the British Bifurcation Coronary Study.;Behan MW;Circulation,2011
2. Double versus single stenting for coronary bifurcation lesions: a meta-analysis.;Katritsis DG;Circulation,2009
3. Bifurcation stenting: current strategies and new devices
4. A randomized clinical study comparing double kissing crush with provisional stenting for treatment of coronary bifurcation lesions results from the DKcrush-II (double kissing crush versus provisional stenting technique for treatment of coronary bifurcation lesions) trial.;Chen SL;J Am Coll Cardiol,2011
5. Bioresorbable Scaffold
Cited by
61 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献