Affiliation:
1. From the Section of Public Health and Health Policy (D.A., R.S., J.P.P.) and Robertson Centre for Biostatistics (A.M.), University of Glasgow; Western Infirmary (K.G.O.); Glasgow Royal Infirmary (H.E.); Victoria Infirmary (R.J.N.), Glasgow; Edinburgh Royal Infirmary (A.D.F.); Western General Hospital (I.R.S.), Edinburgh; Aberdeen Royal Infirmary (K.P.J.), Aberdeen; and Monklands Hospital (A.C.H.P.), Airdrie, Lanarkshire, United Kingdom.
Abstract
Background—
The US Food and Drug Administration recently concluded that data on off-label drug-eluting stent (DES) safety are limited. However, in actual clinical practice, DES are often used for off-label indications, and observational studies demonstrate that complications are higher when compared with on-label use. We aimed to determine whether clinical outcomes differ after DES and bare-metal stent implantation in a patient cohort defined by DES off-label indications.
Methods and Results—
We used the national revascularization registry in Scotland to identify patients who underwent coronary stenting for an off-label indication between January 2003 and September 2005. Individual-level linkage to comprehensive national admission and death databases was used to ascertain the end points of death, myocardial infarction, and target-vessel revascularization. We calculated propensity scores on the basis of clinical, demographic, and angiographic variables and matched DES to bare-metal stents on a 1:1 basis. The final study population consisted of 1642 patients, well matched for important covariables at baseline. Event-free survival was calculated over 24 months with the Kaplan-Meier method. All-cause death was more common after bare-metal stent implantation during follow-up (7.7% versus 6.6%; hazard ratio 0.63; 95% confidence interval, 0.40 to 0.99;
P
=0.04). No difference in the rates of myocardial infarction were noted (7.3% versus 7.5%; hazard ratio 1.02; 95% confidence interval, 0.69 to 1.54;
P
=0.92). Target-vessel revascularization was reduced in patients treated with DES (13.9% versus 10.7%; hazard ratio 0.67; 95% confidence interval, 0.49 to 0.93;
P
=0.02).
Conclusions—
At 24 months, patients treated with DES for off-label indications had lower rates of death and target-vessel revascularization and similar rates of myocardial infarction, as compared with patients treated with bare-metal stents.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Cardiology and Cardiovascular Medicine
Reference37 articles.
1. Sirolimus-Eluting Stents versus Standard Stents in Patients with Stenosis in a Native Coronary Artery
2. One-Year Clinical Results With the Slow-Release, Polymer-Based, Paclitaxel-Eluting TAXUS Stent
3. US Food and Drug Administration. CYPHER Sirolimus-eluting Coronary Stent on RAPTOR Over-the-Wire Delivery System or RAPTORRAIL Rapid Exchange Delivery System - P020026. Available at: http://www.fda.gov/cdrh/pdf2/P020026.html. Accessed January 8 2008.
4. US Food and Drug Administration. TAXUS Express 2 Paclitaxel-Eluting Coronary Stent System (Monorail and Over-the-Wire) - P030025. Available at: http://www.fda.gov/cdrh/pdf3/P030025.html. Accessed January 8 2008.
5. Cause of death with bare metal and sirolimus-eluting stents
Cited by
14 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献