Direct Left Atrial Pressure Monitoring in Ambulatory Heart Failure Patients

Author:

Ritzema Jay1,Melton Iain C.1,Richards A. Mark1,Crozier Ian G.1,Frampton Chris1,Doughty Robert N.1,Whiting James1,Kar Saibal1,Eigler Neal1,Krum Henry1,Abraham William T.1,Troughton Richard W.1

Affiliation:

1. From the Department of Cardiology, Christchurch Hospital, Christchurch, New Zealand (J.R., I.C.M., A.M.R., I.G.C., C.F., R.W.T.); Auckland City Hospital, Auckland, New Zealand (R.N.D.); Cedars-Sinai Medical Center, Los Angeles, Calif (J.W., S.K., N.E.); The Alfred Hospital, Melbourne, Australia (H.K.); and Ohio State University, Columbus (W.T.A.).

Abstract

Background— We describe the first human experience with a permanently implantable, direct left atrial pressure (LAP) monitoring system in ambulatory patients with chronic heart failure. Methods and Results— Eight patients with established heart failure and at least 1 heart failure hospitalization or unplanned visit for parenteral therapy in the last year underwent device implantation under fluoroscopic guidance. All subjects received aspirin 150 mg and clopidogrel 75 mg daily. Subjects measured LAP twice daily and attended a clinic regularly for data upload and device calibration. Right heart catheterization was performed at the time of device implantation and at 12 weeks. The device was implanted in all subjects with no procedural complications. At the 12-week follow-up, 87% of device LAP measurements were within ±5 mm Hg of simultaneous pulmonary capillary wedge pressure readings over a wide range of pressures (1.6 to 71 mm Hg). Net drift corrected by calibration was −0.2±1.9 mm Hg/mo. During short-term follow-up, there were no device-related complications or systemic emboli. There were no deaths, no unplanned heart failure clinic visits, and no admissions for heart failure. Conclusions— Ambulatory monitoring of direct LAP with a new implantable device was well tolerated, feasible, and accurate at a short-term follow-up. Further follow-up and investigation are warranted to evaluate the clinical utility of LAP monitoring in patients with heart failure.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 123 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3