Electromagnetic Interference With Implantable Cardioverter-Defibrillators at Power Frequency

Author:

Napp Andreas1,Joosten Stephan1,Stunder Dominik1,Knackstedt Christian1,Zink Matthias1,Bellmann Barbara1,Marx Nikolaus1,Schauerte Patrick1,Silny Jiri1

Affiliation:

1. From the Department of Internal Medicine I (Cardiology, Pneumology, Angiology) (A.N., C.K., M.Z., B.B., N.M., P.S.) and Research Center for Bioelectromagnetic Interaction at the Institute of Occupational Medicine; former at the Institute of Hygiene and Environmental Medicine (S.J., D.S., J.S.), University Hospital RWTH Aachen, Aachen, Germany; and Department of Cardiology, Maastricht University Medical Center, Maastricht, The Netherlands (C.K.).

Abstract

Background— The number of implantable cardioverter-defibrillators (ICDs) for the prevention of sudden cardiac death is continuing to increase. Given the technological complexity of ICDs, it is of critical importance to identify and control possible harmful electromagnetic interferences between various sources of electromagnetic fields and ICDs in daily life and occupational environments. Methods and Results— Interference thresholds of 110 ICD patients (1-, 2-, and 3-chamber ICDs) were evaluated in a specifically developed test site. Patients were exposed to single and combined electric and magnetic 50-Hz fields with strengths of up to 30 kV·m −1 and 2.55 mT. Tests were conducted considering worst-case conditions, including maximum sensitivity of the device or full inspiration. With devices being programmed to nominal sensitivity, ICDs remained unaffected in 91 patients (83%). Five of 110 devices (5%) showed transient loss of accurate right ventricular sensing, whereas 14 of 31 (45%) of the 2- and 3-chamber devices displayed impaired right atrial sensing. No interference was detected in 71 patients (65%) within the tested limits with programming to maximum sensitivity, whereas 20 of 110 subjects (18%) exhibited right ventricular disturbances and 19 of 31 (61%) subjects exhibited right atrial disturbances. Conclusions— Extremely low-frequency daily-life electromagnetic fields do not disturb sensing capabilities of ICDs. However, strong 50-Hz electromagnetic fields, present in certain occupational environments, may cause inappropriate sensing, potentially leading to false detection of atrial/ventricular arrhythmic events. When the right atrial/right ventricular interferences are compared, the atrial lead is more susceptible to electromagnetic fields. Clinical Trial Registration— URL: http://clinicaltrials.gov/ct2/show/NCT01626261 . Unique identifier: NCT01626261.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Reference39 articles.

1. The Influence of Electromagnetic Environment on the Performance of Artificial Cardiac Pacemakers

2. Institute for Applied Quality Improvement and Research in Health Care GmbH (AQUA). German Hospital Quality Report 2011. http://www.sqg.de/quality-report/index.html. 2012;53–58. Accessed April 4 2013.

3. Inappropriate Implantable Cardioverter-Defibrillator Shocks in MADIT II

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3