Increasing fibrillation duration enhances relative asymmetrical biphasic versus monophasic defibrillator waveform efficacy.

Author:

Jones J L1,Swartz J F1,Jones R E1,Fletcher R1

Affiliation:

1. Department of Physiology and Biophysics, Georgetown University, Washington, DC 20037.

Abstract

Biphasic waveforms reduce defibrillation threshold compared with corresponding monophasic waveforms. However, effects of fibrillation duration on relative efficacy of monophasic and biphasic waveforms are unknown. This study used a newly developed defibrillation model, the isolated right- and left-sided working rabbit heart, with epicardial defibrillation electrodes, to compare threshold for a monophasic waveform (5 msec rectangular) and an asymmetrical biphasic waveform (5 msec each pulse, V2 = 50% V1). Mean voltage defibrillation threshold (V50) was determined from sigmoidal probability of successful defibrillation versus shock intensity curves after 5, 15, and 30 seconds of fibrillation in a paired study with 10 hearts. Results showed that biphasic waveforms had significantly lower voltage and energy thresholds at all fibrillation durations and that their relative efficacy improved with increasing fibrillation duration. Biphasic voltage threshold was 38.2 +/- 2.2, 44.7 +/- 4.8, and 46.6 +/- 3.2 V after 5, 15, and 30 seconds of fibrillation compared with monophasic thresholds of 51.7 +/- 4.4 (p less than 0.002), 63.0 +/- 7.6 (p less than 0.05), and 72.1 +/- 3.9 V (p less than 0.005). Biphasic waveform energy threshold was 0.67 that for the monophasic waveform after 5 seconds of fibrillation (0.12 +/- 0.01 versus 0.18 +/- 0.03 J, p less than 0.05). The ratio between biphasic waveform threshold and monophasic waveform threshold (B/M) decreased to 0.62 at 15 seconds. At 30 seconds, B/M was 0.52 (0.17 +/- 0.02 versus 0.33 +/- 0.04 J, p less than 0.02). This study also showed that biphasic waveform threshold was a nonlinear function of monophasic waveform threshold so that improved biphasic defibrillator waveform efficacy was greatest for hearts having higher monophasic thresholds.(ABSTRACT TRUNCATED AT 250 WORDS)

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Cardiology and Cardiovascular Medicine,Physiology

Cited by 71 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3