Abstract
Instantaneous velocity of pulmonary artery blood was measured using a thin-film resistance anemometer mounted on a hypodermic needle. Studies were performed, at thoracotomy, in five dogs and five patients. Instantaneous blood velocity was recorded at several sites across the transverse axis of the main pulmonary artery to determine the shape of the velocity profile.
We found the velocity profile, in both man and dog, to be approximately flat. In dog, the mean velocity, normalized to that at the center line, ranged from 1.18 (SD ± 0.03) to 0.81 (SD ± 0.27). In man, mean velocity of pulmonary artery blood, normalized in a similar fashion, ranged from 1.08 (SD ± 0.03) to 0.84 (SD ±0.07), while the normalized peak systolic velocity ranged from 1.04 (SD ± 0.07) to 0.95 (SD ± 0.16). The results, in dogs, suggest that there is little asymmetry of the velocity profile and therefore these have been taken as evidence that the use of cuff electromagnetic flowmeters on the pulmonary artery of dogs is not subject to significant inaccuracies associated with a nonsymmetrical flow profile. The finding of a relatively flat velocity profile in patients, moreover, will make it easier for catheter-tip flowmeters to measure bulk velocity in the pulmonary artery.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Cardiology and Cardiovascular Medicine,Physiology
Reference11 articles.
1. Velocity distribution and transition in the arterial system. In Ciba Foundation Symposium on Circulatory and Respiratory Mass Transport, edited by G. E. W. Wolstenholme and Julie Knight. London;SCHULTZ D. L.;J. & A. Churchill,1969
2. TUNSTALL PEDOE D. S.: Velocity distribution of blood flow in major arteries of animals and man. D. Phil. Thesis University of Oxford 1970.
3. Problems in the Measurement of Blood Flow by Magnetic Induction
4. GOLDMAN S. C. MARPLE N. B. AND SCOLNICK W. L.: Effects of flow profile on electromagnetic flowmeter accuracy. J Appl Physio! 18: 652 1963.
5. BERCEL D. H. AND GESSNER U.: The Electro
Cited by
36 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献