Effect of Hematocrit Value upon Electromagnetic Flowmeter Sensitivity

Author:

DENNIS JENNIFER1,WYATT D. G.1

Affiliation:

1. Nuffield Institute for Medical Research, University of Oxford, Oxford, England

Abstract

An electromagnetic flowmeter system that has an accuracy of 0.1% and is insensitive to fluid conductivity was used to study the sensitivity to steady flow of a 6-mm cannular transducer. All measurements were carried out at 37°C. Sensitivity decreased with increasing hematocrit value whether flow was laminar or otherwise. With laminar flow and a given hematocrit value, the sensitivity also decreased with increasing flow rate. Sensitivity started to rise when increasing flow rate caused departure from the laminar regime. At the lowest flow used in the laminar regime (250 ml/min), the sensitivity fell 4% when the hematocrit value was changed from 0 to 66% at 39.5% the reduction in sensitivity was 2%. With laminar flow at hematocrit 39.5% the sensitivity fell 1% as flow increased from 250 to 1500 ml/min. At hematocrit 29.5% the sensitivity rose 1.2% when the flow regime changed from laminar to turbulent. The changes may be explained in terms of a radial distribution of cells, of which a cell-free boundary layer is a particular example. Theoretical calculations show that the thickness of an apparent cell-free boundary layer must be tens of microns to explain the observations with laminar flow and that the apparent thickness is greatly reduced with onset of turbulence. A possible mechanism is described by which the results could be explained in terms of cell orientation.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Cardiology and Cardiovascular Medicine,Physiology

Reference27 articles.

1. SHERCLIFF J. A.: Theory of electromagnetic flow measurement. Cambridge Cambridge University Press 1962.

2. Baseline errors in cuff electromagnetic flowmeters

3. Noise in electromagnetic flowmeters

4. Electromagnetic flowmeter for use with intact vessels. J. Physiol;WYATT D. G.;London,1964

5. An input transformer with low earth leakage currents;WYATT D. G.;Electron. Eng,1965

Cited by 31 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Use of Fluorescence Guidance in Cardiothoracic Surgery;The SAGES Manual of Fluorescence-Guided Surgery;2023

2. Angiography During Cardiovascular Surgery;Video Atlas of Intraoperative Applications of Near Infrared Fluorescence Imaging;2020

3. Design and Optimization of Annular Flow Electromagnetic Measurement System for Drilling Engineering;Journal of Sensors;2018

4. Comparisons between electromagnetic and X-beam transit-time flow measurements for evaluating drug actions on cardiac output in the conscious dog;Journal of Pharmacological and Toxicological Methods;2006-11

5. Digital tools to facilitate intraoperative coronary artery bypass graft patency assessment;Seminars in Thoracic and Cardiovascular Surgery;2004-09

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3