1. From the Hypertension Center STRIDE-7, National and Kapodistrian University of Athens, School of Medicine, Third Department of Medicine, Sotiria Hospital, Athens, Greece (G.S.S.); University of Tennessee Health Science Center (B.A.); Physikalisch-Technische Bundesanstalt, Berlin, Germany (S.M.); Foundation, Medical Research Institutes, Paris, France (R.A.); Medaval, Dublin, Ireland (N.A.); Clinic for Cardiology, Herz- und Diabeteszentrum NRW, Ruhr-Universität Bochum, Bad Oeynhausen, Germany (S.E.);...
Abstract
In the past 30 years, several organizations, such as the US Association for the Advancement of Medical Instrumentation (AAMI), the British Hypertension Society, the European Society of Hypertension (ESH) Working Group on Blood Pressure (BP) Monitoring, and the International Organization for Standardization (ISO), have developed protocols for clinical validation of BP measuring devices. However, it is recognized that science, as well as patients, consumers, and manufacturers, would be best served if all BP measuring devices were assessed for accuracy according to an agreed single validation protocol that had global acceptance. Therefore, an international initiative was taken by the AAMI, ESH, and ISO experts who agreed to develop a universal standard for device validation. This statement presents the key aspects of a validation procedure, which were agreed by the AAMI, ESH, and ISO representatives as the basis for a single universal validation protocol. As soon as the AAMI/ESH/ISO standard is fully developed, this will be regarded as the single universal standard and will replace all other previous standards/protocols.