Effect of prosthetic valve malfunction on the Doppler-catheter gradient relation for bileaflet aortic valve prostheses.

Author:

Baumgartner H1,Schima H1,Kühn P1

Affiliation:

1. Second Department of Internal Medicine/Cardiology, Krankenhaus der Barmherzigen Schwestern, Linz, Austria.

Abstract

BACKGROUND Considerable discrepancies between Doppler and catheter gradients caused by localized gradients and pressure recovery have been reported for normal bileaflet aortic valve prostheses. METHODS AND RESULTS To examine whether this Doppler-catheter gradient relation is affected by prosthetic valve malfunction, a 19-mm CarboMedics aortic valve was simultaneously studied with continuous-wave Doppler and catheter technique in normal function and in various states of malfunction ranging from slightly restricted opening to total occlusion of one leaflet. For each functional status, peak and mean gradients were measured at eight different flow rates (cardiac output, 2.0-6.0 L/min). Excellent correlation between Doppler and catheter gradients was found regardless of the valve function (r = 0.99, SEE = 1.0-3.3 mm Hg). However, the relation between Doppler and catheter gradient was highly dependent on the function of the valve as shown by a variation of slopes from 1.08 to 2.08. For the normally functioning valve (angle between flow axis and leaflet 5 degrees), peak and mean Doppler gradients were approximately twice the catheter gradients (slope, 2.08 and 2.03 for peak and mean gradients, respectively). Slightly restricted opening of one leaflet (22 degrees) significantly altered the Doppler-catheter gradient relation, and slopes decreased to 1.69 (p < 0.01) and 1.52 (p < 0.001) for peak and mean gradients, respectively. The differences between Doppler and catheter gradients significantly decreased with further restriction of valve opening, and slopes ranged from 1.25 to 1.41 for angles between 34 degrees and 52 degrees. When one leaflet was totally occluded, the slope finally dropped to 1.08 for both peak and mean gradients, and Doppler gradients were only slightly greater than catheter gradients. Gradients increased with malfunction of the valve caused by reduction of the effective orifice area. However, the increase of Doppler gradients was considerably smaller than the increase of simultaneous catheter gradients. CONCLUSIONS The discrepancies between Doppler and catheter gradients that have been reported for normally functioning bileaflet aortic valve prostheses may be reduced or even disappear in patients with malfunctioning valves. Furthermore, the increase of Doppler gradients caused by malfunction of the valve may underestimate the true hemodynamic changes.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3