Prospective, randomized comparison in humans of a unipolar defibrillation system with that using an additional superior vena cava electrode.

Author:

Bardy G H1,Dolack G L1,Kudenchuk P J1,Poole J E1,Mehra R1,Johnson G1

Affiliation:

1. Department of Medicine, University of Washington, Seattle.

Abstract

BACKGROUND A unipolar defibrillation system using a single right ventricular (RV) electrode and the active shell or "CAN" of the implantable cardioverter-defibrillator itself situated in a left infraclavicular pocket has been shown to be as efficient in defibrillation as an epicardial lead system. The purpose of this study was to determine whether defibrillation efficacy can be improved further by adding a superior vena cava (SVC) electrode to this already efficient defibrillation system. METHODS AND RESULTS We prospectively and randomly compared the defibrillation efficacy of a simplified unipolar defibrillation system, RV-->CAN, with that of one incorporating a high SVC electrode, RV-->SVC + CAN, in 15 consecutive cardiac arrest survivors undergoing implantation of a presently available transvenous defibrillation system. The RV defibrillation electrode was a 5-cm coil located on a 10.5F lead used as the anode in both lead configurations examined. The active CAN was a 108-cm2 surface area shell of a titanium alloy pulse generator used as the cathode in both configurations and placed in a left infraclavicular pocket. The SVC electrode was a 6F 5-cm-long coil and was used as an additional cathode positioned at the junction of the SVC and the left innominate vein. The defibrillation pulse used was a 65% tilt, asymmetric biphasic waveform delivered from a 120-microF capacitor. The defibrillation threshold (DFT) stored energy, leading edge voltage, current, and pulsing resistance were measured for both lead systems. The single-lead unipolar system, RV-->CAN, resulted in a stored energy DFT of 7.4 +/- 5.2 J, and the three-electrode dual pathway system, RV-->SVC + CAN, resulted in a DFT of 6.0 +/- 3.4 J (P = .20). There was no difference in defibrillation efficacy with the more complicated three-electrode system over the unipolar system despite a decrease in pulsing resistance to 48.6 +/- 3.5 omega compared with 61.2 +/- 5.9 omega for the unipolar system (P < .0001) and a slight rise in delivered current to 6.3 +/- 1.8 A compared with 5.5 +/- 2.0 A for the unipolar system (P = .062). CONCLUSIONS The unipolar single-lead transvenous defibrillation system provides defibrillation at energy levels comparable to that reported with present epicardial lead systems. Coupling of this lead system to a third SVC electrode increases system complexity but offers little defibrillation advantage despite a large decrease in pulsing resistance and a modest increase in delivered current.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 89 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3