Catheter ablation of atrioventricular junction using radiofrequency current in 17 patients. Comparison of standard and large-tip catheter electrodes.

Author:

Jackman W M1,Wang X Z1,Friday K J1,Fitzgerald D M1,Roman C1,Moulton K1,Margolis P D1,Bowman A J1,Kuck K H1,Naccarelli G V1

Affiliation:

1. Department of Medicine, University of Oklahoma Health Sciences Center, Oklahoma City 73190.

Abstract

BACKGROUND Two catheter electrode systems were compared for delivering radiofrequency current for ablation of the atrioventricular junction. Seventeen patients with drug-resistant supraventricular tachyarrhythmias were studied. METHODS AND RESULTS A 6F or 7F catheter with six or eight standard electrodes (1.25 mm wide, 2.5-mm spacing) was used in the first seven patients (group 1). A 7F quadripolar catheter with a large-tip electrode (4 mm long; surface area, 27 mm2) was used in the final 10 patients (group 2). Both ablation catheters were positioned to record a large atrial potential and a small but sharp His bundle potential from the distal bipolar electrode pair. Radiofrequency current was applied between a large skin electrode on the left posterior chest and either 1) each individual electrode on the standard-tip electrode catheter at 40 V (group 1) or 2) the large-tip electrode at 50-60 V (group 2). Radiofrequency current was limited to 40 V in group patients because of the strong potential for an early impedance rise when higher voltage is applied through standard electrodes. Complete atrioventricular block was achieved in six of seven group 1 patients and all 10 group 2 patients. A junctional escape rhythm followed ablation in five or six group 1 patients (mean cycle length, 1,066 +/- 162 msec) and eight of 10 group 2 patients (mean cycle length, 1,281 +/- 231 msec). Atrioventricular block was produced in a mean of 4.7 +/- 4.6 radiofrequency current applications delivered over a period of 42 +/- 45 minutes using the large-tip electrode (group 2) compared with 46 +/- 22 applications using standard electrodes (15.9 +/- 10.2 applications delivered through the standard-tip electrode) over a period of 147 +/- 59 minutes (group 1). For the application producing atrioventricular block, the large-tip electrode used higher voltage (58 +/- 17 versus 38 +/- 5 V, p less than 0.03) and had lower impedance (103 +/- 22 versus 148 +/- 40 omega, p less than 0.01), resulting in greater power (33.0 +/- 13.0 versus 10.2 +/- 0.6 W, p less than 0.003) and shorter time to block (8 +/- 3 versus 22 +/- 3 seconds, p less than 0.001). Current delivery through standard electrodes was limited by an impedance rise occurring 7 +/- 7 seconds after the onset of one or more radiofrequency current applications at 10 +/- 1 W in six of seven patients. Using the large-tip electrode, an impedance rise occurred in five of 10 patients, but at 25 +/- 10 W and after 21 +/- 9 seconds. Atrioventricular block occurred before the impedance rise in three of these five patients. Complete atrioventricular block persisted in 15 of 16 patients at a mean follow-up of 8.7 months. Atrioventricular conduction returned at 1 month in one group 2 patient and was successfully ablated by a second procedure. Three group 1 patients died 0.5-2 months after ablation, and a fourth patient underwent cardiac transplantation after 10 months. Pathological examination of the heart in two of these patients showed necrosis of the atrioventricular node and origin of the His bundle, without injury to the middle or distal His bundle. All 10 group 2 patients are alive and subjectively improved after ablation. CONCLUSIONS We conclude that catheter-delivered radiofrequency current effectively produces complete atrioventricular block (94%) without requiring general anesthesia or the risk of ventricular dysfunction or cardiac perforation. The large-tip electrode allows a threefold increase in delivered power and markedly decreases the number of pulses and time required to produce atrioventricular block.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 171 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Anatomical Ablation of the Atrioventricular Node;Arrhythmia & Electrophysiology Review;2024-08-20

2. Surgical and Transcatheter Management of Arrhythmias;Pediatric Cardiac Surgery;2023-02-03

3. Accessory pathway recording and ablation;Heart Rhythm;2021-05

4. The early translational research of radiofrequency catheter ablation;Heart Rhythm;2021-02

5. Emerging Technologies for Pulmonary Vein Isolation;Circulation Research;2020-06-19

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3