DeCaprio V,Hurzeler P,Furman S
Abstract
Simultaneous unipolar and bipolar electrograms were recorded and compared from 49 pacemaker patients with bipolar endocardial electrodes. Average bipolar depolarization signal voltage equalled that of unipolar but showed greater variation. Bipolar and unipolar slew rates were equal in both mean and variance. The proximal pole voltage had little effect on the bipolar result in 8% of the cases, tended to cancel the tip voltage in 49% of the cases and augmented the tip voltage in 43% of the electrograms. The average bipolar R wave duration was 28% less, the T wave amplitude 34% less, and the ST-segment elevation 37% less than the unipolar values. By consistently attenuating the undersirable T waves and ST elevations, while leaving the depolarization signal unaffected, the bipolar electrode offered the advantage of a superior signal-to-noise ratio for sensing depolarization. In one case, however, the bipolar signal was so small as to cause a clinical sensing failure.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Physiology (medical),Cardiology and Cardiovascular Medicine
Reference14 articles.
1. Lewis T: The Mechanism and Graphic Registration of the Heart Beat. London Shaw and Sons Ltd 1925
2. Barold SS Keller JE: Sensing problems with demand pacemakers. In Cardiac Pacing edited by Samet P. New York Grune and Stratton 1973
3. Failure of Demand Pacemaker From Low-Voltage Bipolar Ventricular Electrograms
4. In Technical and clinical aspects of pacing electrodes;Greatbatch W;Ann NY Acad Sci,1969
5. Endocardial electrograms and pacemaker sensing;Hurzeler P;Med Instrum,1976
Cited by
89 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献