Statistical analysis of pacemaker follow-up data. Rate stability and reliability.

Author:

Grunkemeier G L,Dobbs J L,Starr A

Abstract

In a rate-stable pacemaker, pacemaker rate reflects battery voltage level and should remain constant or be slightly depleted with time in a linear fashion until failure occurs. Linear regression techniques are therefore particularly suited to the analysis of pacemaker rate data. Data on model 8114/8114A Starr-Edwards pacemakers followed at the University of Oregon Health Sciences Center were studied. For a given pacemaker both the slope of the regression line and the standard deviation about the regression line are proportional to the original rate. Natural indices of rate depletion and rate stability for each pacemaker can therefore be calculated by dividing the slope and the standard deviation, respectively, by the inpercept of the corresponding regression line. This model pacemaker has a small average standard deviation so that significant variations from the line of regression, indicating impending pulse generator failure, can be detected in spite of random fluctuations. Emergency and prophylactic replacements can therefore be kept to a minimum so that pulse generator lifetime is maximized. Success in extending pulse generator longevity can be measured by actuarial techniques, which show this pacemaker to have a median battery depletion time of 38 months and a median replacement time of 35 months.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 14 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Electroactive Nanomaterials for the Prevention and Treatment of Heart Failure: From Materials and Mechanisms to Applications;Small;2024-09-13

2. Klinischer Teil;Langzeit-Elektrokardiographie;1996

3. Klinischer Teil;Langzeit-Elektrokardiographie;1992

4. Evaluation of pacemaker pulse generator and patient longevity in patients aged 1 day to 20 years;The American Journal of Cardiology;1987-04

5. Pacemaker Therapy;Surgical Clinics of North America;1985-06

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3