A chronically implanted system for automatic defibrillation in active conscious dogs. Experimental model for treatment of sudden death from ventricular fibrillation.
-
Published:1978-07
Issue:1
Volume:58
Page:90-94
-
ISSN:0009-7322
-
Container-title:Circulation
-
language:en
-
Short-container-title:Circulation
Author:
Mirowski M,Mower M M,Langer A,Heilman M S,Schreibman J
Abstract
Ventricular defibrillation was acheived in active conscious dogs with a chronically implanted automatic system composed of a defibrillator and an alternating current fibrillator. The hermetically sealed defibrillator is encased in titanium, weighs 250 g and has a volume of 145 ml. The sensor continuously monitors ventricular electircal activity and recognizes fibrillation by the absence of isoelectric potential segments. Fibrillation is induced by placing a magnet over the implanted fibrillator. The resulting syncope closely resembles the clinical entity of sudden death, while the defibrillator automatically restores normal rhythm with a truncated exponential pulse of 30 J, 15 seconds after the onset of the arrhythmia. The operational status of the defibrillator can be tested in vitro and noninvasively in vivo with an external analyzer. This experimental model allows for the first time a long-term study of the automatic implantable defibrillator approach to prevent sudden death from ventricular fibrillation under a variety of physiopathologic conditions.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Physiology (medical),Cardiology and Cardiovascular Medicine
Reference33 articles.
1. Standby Automatic Defibrillator
2. Experimental ventricular defibrillation with an automatic and completely implanted system;Schuder JC;Trans Am Soc Artif Int Organs,1970
3. Heilman MS Langer A Mower MM Mirowski M: Analysis of four implantable electrode systems for automatic defibrillator. (abstr) Circulation 52 (suppl II): II-194 1975
4. Considerations in the development of the automatic implantable defibrillator;Langer A;Med Instrum,1976
5. Transthoracic Ventricular Defibrillation with Triangular and Trapezoidal Waveforms
Cited by
117 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献