Permanent Ventricular Pacemakers

Author:

MORRIS JAMES J.1,WHALEN ROBERT E.1,MCINTOSH HENRY D.1,THOMPSON HOWARD K.1,BROWN IVAN W.1,YOUNG W. GLENN1

Affiliation:

1. From the Cardiovascular Laboratory, Department of Medicine, and the Department of Surgery, Duke University Medical Center, Durham, North Carolina.

Abstract

In the series of 86 patients with the Stokes-Adams syndrome or with symptomatic bradycardia managed with permanently implanted ventricular pacemakers reported on, 35 patients had primary implantation of epicardial leads at thoracotomy and 51 had transvenous endocardial electrodes passed via the jugular venous system for permanent ventricular pacing. The transvenous method of permanent pacemaker implantation appears to be easier to apply with less serious complications and provides the same overall mortality as the transthoracic approach. However, pacemaker failure, both permanent and temporary, is more likely to occur with the transvenous approach. A series of unique complications and findings noted in the transvenous group were thought to be related to the permanent pacing electrode catheter. These included myothought cardial perforation, pericardial friction rubs, intermittent ventricular pacing, and diaphragmatic stimulation. The experience suggests that it is reasonable to employ a transvenous pacemaker initially, recognizing that replacement with a transthoracic pacemaker will be necessary in approximately 16% of the patients and remanipulation of the electrode catheter will be needed in another 13%. Therefore, to undertake transvenous permanent implantation of a pacemaker, one must also be prepared to employ the transthoracic approach in a small but significant number of patients.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Reference29 articles.

1. Transistorized, self-contained, implantable pacemaker for the long-term correction of complete heart block;CHARDACK W. M.;Surgery,1960

2. Long-Term Electric Stimulation of the Heart for Stokes-Adams Disease

3. Treatment of complete heart block with an implanted controllable pacemaker;KANTROWITZ A.;Surg Gynec Obstet,1962

4. New technique for internal cardiac pacing;SMDONS A. H.;Lancet,1963

5. Dipolar Electrode in Heart Block

Cited by 57 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Malplacement of Endocardial Pacemaker Electrodes in the middle cardiac vein;Acta Medica Scandinavica;2009-04-24

2. Experiences with a new myocardial electrode for permanent cardiac pacing;Acta Medica Scandinavica;2009-04-24

3. Displacement of endocardial pacemaker electrodes;Acta Medica Scandinavica;2009-04-24

4. PERMANENT ENDOCARDIAL PACING;Acta Medica Scandinavica;2009-04-24

5. Emergency cardiac pacing;The American Journal of Emergency Medicine;1993-03