Accurate Assessment of Patient Effective Radiation Dose and Associated Detriment Risk From Radiofrequency Catheter Ablation Procedures

Author:

Perisinakis Kostas1,Damilakis John1,Theocharopoulos Nicholas1,Manios Emmanouel1,Vardas Panayiotis1,Gourtsoyiannis Nicholas1

Affiliation:

1. From the University of Crete, Faculty of Medicine, Iraklion, Crete, Greece.

Abstract

Background— Radiofrequency (RF) cardiac catheter ablation procedures may require extended fluoroscopic exposure resulting in elevated radiation risk. The aim of the present study was to accurately establish RF ablation radiation risk levels and to provide means for accurate patient risk estimation from studies performed in any electrophysiology laboratory. Methods and Results— Fluoroscopy required during cardiac ablation was classified into 4 types identified by beam orientation and irradiated tissue: (1) posteroanterior exposure during catheter advancing from the groin to the heart, (2) posteroanterior heart exposure, (3) left anterior oblique heart exposure, and (4) right anterior oblique heart exposure. The duration of each exposure was monitored in 24 patients undergoing RF cardiac ablation. Dose per minute of fluoroscopy was measured at 15 organs/tissues for each projection with the use of anthropomorphic phantom and thermoluminescence dosimetry. The effective dose rate was 219, 144, 136, and 112 μGy/min for groin-to-heart posteroanterior, posteroanterior, left anterior oblique, and right anterior oblique exposure, respectively. A typical ablation procedure results in a total effective dose of 8.3 mSv per hour of fluoroscopy. The average excess of fatal cancers was estimated to be 650 and 480 per million patients undergoing RF ablation requiring 1 hour of fluoroscopy for US and UK populations, respectively. The average risk for genetic defects was determined to be 1 per million births. Conclusions— Radiation risk from RF cardiac ablation is moderate compared with other complications, but it may highly exceed radiation risk from common radiological procedures. Efforts should be made toward minimization of patient radiation risk from RF ablation procedures.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3