Validation Study to Determine the Accuracy of Central Blood Pressure Measurement Using the Sphygmocor Xcel Cuff Device

Author:

Schultz Martin G.1,Picone Dean S.1,Armstrong Matthew K.1,Black J. Andrew12,Dwyer Nathan2,Roberts-Thomson Philip12,Sharman James E.1

Affiliation:

1. From the Menzies Institute for Medical Research, University of Tasmania, Hobart, Australia (M.G.S., D.S.P., M.K.A., J.A.B., N.D., P.R.-T., J.E.S.)

2. Royal Hobart Hospital, Hobart, Australia (J.A.B., N.D., P.R.-T.).

Abstract

Numerous devices purport to measure central (aortic) blood pressure (BP) as distinct from conventional brachial BP. This validation study aimed to determine the accuracy of the Sphygmocor Xcel cuff device (AtCor Medical, CardieX, Sydney, Australia) for measuring central BP. 296 patients (mean age 61±12 years) undergoing coronary angiography had simultaneous measurement of invasive central BP and noninvasive cuff-derived central BP using the Xcel cuff device (total n=558 individual comparisons). A subsample (n=151) also had invasive brachial BP measured. Methods were undertaken according to the Artery Society recommendations, and several calibration techniques to derive central systolic BP (SBP) were examined. Minimum acceptable error was ≤5±≤8 mm Hg. Central SBP was significantly underestimated, and with wide variability, when using the default calibration of brachial-cuff SBP and diastolic BP (DBP; mean difference±SD, −7.7±11.0 mm Hg). Similar variability was observed using other calibration methods (cuff 33% form-factor mean arterial pressure and DBP, −4.4±11.5 mm Hg; cuff 40% form-factor mean arterial pressure and DBP, 4.7±11.9 mm Hg; cuff oscillometric mean arterial pressure and DBP, −18.2±12.1 mm Hg). Only calibration with invasive central integrated mean arterial pressure and DBP was within minimal acceptable error (3.3±7.5 mm Hg). The difference between brachial-cuff SBP and invasive central SBP was 3.3±10.7 mm Hg. A subsample analysis to determine the accuracy of central-to-brachial SBP amplification showed this to be overestimated by the Xcel cuff device (mean difference 4.3±9.1 mm Hg, P =0.02). Irrespective of cuff calibration technique, the Sphygmocor Xcel cuff device does not meet the Artery Society accuracy criteria for noninvasive measurement of central BP.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Internal Medicine

Cited by 36 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3