Truncated Biphasic Pulses for Transthoracic Defibrillation

Author:

Bardy Gust H.1,Gliner Bradford E.1,Kudenchuk Peter J.1,Poole Jeanne E.1,Dolack G. Lee1,Jones Gregory K.1,Anderson Jill1,Troutman Charles1,Johnson George1

Affiliation:

1. From the Department of Medicine, University of Washington, Seattle.

Abstract

Background Early defibrillation is the single most important factor for improving out-of-hospital ventricular fibrillation resuscitation rates. To achieve the earlier response times required for survival, typically <6 minutes from time of collapse, it will be necessary to equip a far wider network of first responders (firefighters, police, and other individuals with responsibility for public safety) with small, lightweight, and inexpensive automatic external defibrillators (AEDs). An important step in reducing the size and cost of AEDs will be to improve defibrillation efficacy. Because biphasic waveform defibrillation has had a favorable impact on implantable cardioverter-defibrillators (ICDs), there are reasons to believe that biphasic waveforms would also improve transthoracic defibrillators. Our purpose, therefore, was to examine the efficacy of two different low-energy biphasic truncated waveforms referenced to a standard damped sine waveform for transthoracic defibrillation in humans. Methods and Results We prospectively and randomly compared the transthoracic defibrillation efficacy of two different truncated biphasic waveforms, 115 J (70 μF) and 130 J (105 μF), with that of a standard 200-J (36-μF, 28-mH) damped sine wave pulse using right anterior and left lateral thoracic pads (R2 Medical Systems) in 30 cardiac arrest survivors during transvenous ICD surgery. The right anterior patch electrode was used as the cathode and the left lateral thoracic pad as the anode. Transthoracic ventricular defibrillation rescue shocks were tested after a failed transvenous defibrillation shock delivered in the course of ICD testing. Each of the three different rescue shocks was tested in random order in each patient. All shocks were delivered at end expiration. The investigators responsible for determining transthoracic shock efficacy were blinded throughout the study to the transthoracic rescue waveform used. A total of 33 patients were considered for study, but three patients failed to satisfy all entry criteria or did not have a sufficient number of ventricular fibrillation inductions to allow for testing of all three waveforms. Percent efficacy for the three waveforms was then compared in the 30 patients who satisfied entry criteria and completed the protocol. The study population had a mean age of 61±11 years, with 22 (73%) being men. The mean left ventricular ejection fraction was 0.39±0.14. Coronary artery disease was present in 22 (73%). The 115-J (70-μF) biphasic pulse, the 130-J (105-μF) biphasic pulse, and the 200-J (36-μF, 28-mH) damped sine wave pulse were equally effective, resulting in a 97% first-shock ventricular defibrillation efficacy rate. Each waveform failed to defibrillate once, with each waveform failing in a different patient. Conclusions The results of this study suggest that biphasic truncated transthoracic shocks of low energy (115 and 130 J) are as effective as 200-J damped sine wave shocks used in standard transthoracic defibrillators. This finding may contribute significantly to the miniaturization and cost reduction of transthoracic defibrillators, which could enable the development of a new generation of AEDs appropriate for an expanded group of out-of-hospital first responders and, eventually, layperson use.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3