Abstract
Background
Myocardial uptake of either
201
Tl or
99m
Tc-sestamibi (sestamibi) is known to plateau at high coronary flow rates. However, few direct comparisons have been made between these tracers to determine what effect differences in the uptake plateau for the two tracers may have on the detection of coronary stenoses of various severities.
Methods and Results
Twenty-two dogs were instrumented with flow transducers on the left anterior descending (LAD) and circumflex (LCx) arteries. In 6 nonstenotic dogs, adenosine was infused directly into the LAD, whereas 16 dogs with either critical (n=7) or mild (n=9) LAD stenoses received an intravenous infusion. At peak flow,
201
Tl (0.5 mCi), sestamibi (5 to 8 mCi), and radiolabeled microspheres were injected simultaneously. Five minutes later, dogs were killed, and ex vivo imaging of heart slices and gamma-well counting of multiple myocardial samples was performed. Neither
201
Tl nor sestamibi uptake increased in direct proportion to flow. In the 6 nonstenotic dogs, a fivefold increase in LAD flow increased
201
Tl and sestamibi uptake by only 202±6% and 138±4%, respectively (
P
<.0001). In the dogs with critical stenosis, the ratios of stenotic to normal activity by well counting for
201
Tl (0.37±0.05) and sestamibi (0.53±0.06) underestimated the microsphere-determined flow disparity (0.17±0.03) (
P
<.005), but the degree of underestimation was greater for sestamibi (
P
=.001). Similarly, in the dogs with mild stenosis, the stenotic-to-normal ratio for
201
Tl (0.62±0.04) approximated the flow ratio (0.43±0.04) better than sestamibi (0.79±0.03) (
P
<.0001). Sestamibi defects, however, were visually identifiable on the images of the myocardial slices. By image quantification, sestamibi defect magnitude (LAD-to-LCx count ratio) in the critical stenosis group (0.62±0.05) was significantly less than in the mild stenosis group (0.80±0.02) (
P
<.01).
Conclusions
Thus, with adenosine-induced hyperemic flow, both
201
Tl and sestamibi significantly underestimated the magnitude of the flow disparity between stenotic and normal perfusion beds. The degree of underestimation was greater for sestamibi. The clinical implication of these experimental findings for vasodilator perfusion imaging remains to be determined, since factors such as greater redistribution and scatter with
201
Tl could offset its advantages.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Physiology (medical),Cardiology and Cardiovascular Medicine