Probability of Successful Defibrillation at Multiples of the Defibrillation Energy Requirement in Patients With an Implantable Defibrillator

Author:

Strickberger S. Adam1,Daoud Emile G.1,Davidson Theresa1,Weiss Raul1,Bogun Frank1,Knight Bradley P.1,Bahu Marwan1,Goyal Rajiva1,Man K. Ching1,Morady Fred1

Affiliation:

1. From the Department of Internal Medicine, University of Michigan Medical Ctr, Ann Arbor, Mich.

Abstract

Background The probability of successful defibrillation has been determined in normal animals but not in patients undergoing defibrillator implantation. Therefore, the purpose of this prospective study was to determine the probability of successful defibrillation in humans on the basis of a step-down defibrillation energy requirement. Methods and Results Fifty-three consecutive patients underwent five separate inductions of ventricular fibrillation after the defibrillation energy requirement was determined with the use of small decrements and a step-down protocol (20, 15, 12, 10, 8, 6, 5, 4, 3, 2, 1, and 0.8 J). The first shock energy for defibrillation was either 1.0, 1.3, 1.5, 1.7, or 2.0 times the defibrillation energy requirement, and the likelihoods of successful defibrillation were 70±27%, 84±12%, 86±25%, 80±29%, and 88±32%, respectively ( P =.03). The frequencies of uniformly successful defibrillation (5 of 5 defibrillation attempts) were 30%, 27%, 60%, 64%, and 73%, respectively ( P =.01). Seven patients in whom the defibrillation energy requirement was <4 J had an overall rate of successful defibrillation of 54±20% compared with 86±20% in the remaining 47 patients ( P =.002). The likelihood of successful defibrillation at twice the defibrillation energy requirement was 98% in the 46 patients with a defibrillation energy requirement of >4 J and 67% in the 7 patients with a defibrillation energy requirement of <4 J ( P =.17). An absolute safety margin of 7 J was associated with a 96% probability of successful defibrillation. Conclusions The probability of successful defibrillation is 70% at the defibrillation energy requirement. The probability plateaus at 88%, at twice the defibrillation energy requirement. A 96% probability of successful defibrillation is achieved at an absolute safety margin of 7 J, and a 98% success rate is achieved at energies that are twice the defibrillation energy requirement if the defibrillation energy requirement is >4 J. If the defibrillation energy requirement is <4 J, larger multiples of the defibrillation energy requirement are needed to achieve a high probability of successful defibrillation.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 56 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. The “Defibrillation Testing, Why Not?” survey. Testing of subcutaneous and transvenous defibrillators in the Italian clinical practice;IJC Heart & Vasculature;2022-02

2. Back to the Future;JACC: Clinical Electrophysiology;2021-06

3. Pacing and Defibrillation;Cardiac Pacing, Defibrillation and Resynchronization;2021-04-19

4. The Saga of Defibrillation Testing: When Less Is More;Current Cardiology Reports;2018-05-05

5. Systematic Review of Defibrillation Threshold Testing at De Novo Implantation;Circulation: Arrhythmia and Electrophysiology;2016-04