Limitations and Late Complications of Third-Generation Automatic Cardioverter-Defibrillators

Author:

Nunain S. O.1,Roelke M.1,Trouton T.1,Osswald S.1,Kim Y. H.1,Sosa-Suarez G.1,Brooks D. R.1,McGovern B.1,Guy M.1,Torchiana D. F.1,Vlahakes G. J.1,Garan H.1,Ruskin J. N.1

Affiliation:

1. From the Cardiac Arrhythmia Service and the Cardiac Surgery Service, Harvard Medical School and the Massachusetts General Hospital, Boston.

Abstract

Background This study examines the limitations and complex management problems associated with the use of tiered-therapy, implantable cardioverter-defibrillators (ICDs). Methods and Results The study group comprises the first 154 patients undergoing implantation of tiered-therapy ICDs at our institution. Pulse generators from three different manufacturers were used. In 39 patients, a complete nonthoracotomy lead system was used. The perioperative mortality was 1.3%. Of these 154 patients, 37% experienced late postoperative problems. Twenty-one patients required system revision within 36.5 months (mean, 8.57±11.3) of surgery. Reasons for revision were spurious shocks due to electrode fractures (3) or electrode adapter malfunction (2), inadequate signal from endocardial rate-sensing electrodes (3), superior vena cava or right ventricular coil migration (5), failure to correct tachyarrhythmias due to a postimplant rise in defibrillation threshold (5), or pulse generator failure (3). One of these patients required system removal for infection after revision of an endocardial lead. A further 32 patients received inappropriate shocks for atrial fibrillation with a rapid ventricular response or sinus tachycardia. Two of these patients also received shocks for ventricular tachycardia initiated by antitachycardia pacing triggered by atrial fibrillation. Ventricular pacing for bradycardia was associated with inappropriate shocks due to excessive autogain in 2 patients. Conclusions Despite the major diagnostic and therapeutic advantages of tiered-therapy ICDs, a significant proportion of patients continue to experience hardware-related complications or receive inappropriate shocks.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 132 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. The clinical impact of untreated slow ventricular tachycardia in patients carrying implantable cardiac defibrillators;Journal of Interventional Cardiac Electrophysiology;2020-09-23

2. Device Therapy in the Heart Failure;Heart Failure;2019

3. Value of Tei Index in Predicting Implantable Cardioverter Defibrillator Shocks;SiSli Etfal Hastanesi Tip Bulteni / The Medical Bulletin of Sisli Hospital;2018

4. Inappropriate shock for myopotential over-sensing in a patient with subcutaneous ICD;Indian Heart Journal;2015-01

5. Management of Pacemaker 2;Pediatric Cardiology and Cardiac Surgery;2014

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3