Implantation by electrophysiologists of 100 consecutive cardioverter defibrillators with nonthoracotomy lead systems.

Author:

Strickberger S A1,Hummel J D1,Daoud E1,Niebauer M1,Williamson B D1,Man K C1,Horwood L1,Schmittou A1,Kalbfleisch S J1,Langberg J J1

Affiliation:

1. Department of Internal Medicine, University of Michigan Medical Center, Ann Arbor 18109-0022.

Abstract

BACKGROUND Traditional lead systems for implantable cardioverter defibrillators (ICDs) require a thoracotomy for placement. Nonthoracotomy lead systems are available and are usually implanted by an electrophysiologist and a surgeon. The purpose of this study was to prospectively evaluate the safety and efficacy of ICD implantation with a nonthoracotomy lead system by electrophysiologists. METHODS AND RESULTS A consecutive series of 100 patients (mean age, 61 +/- 13 years, +/- SD) underwent ICD implantation with a nonthoracotomy lead system while intubated and under general anesthesia. Seventy-seven patients had coronary artery disease, 15 had idiopathic cardiomyopathy, 6 had miscellaneous heart disease, and 2 had structurally normal hearts. The mean ejection fraction was 0.29 +/- 0.13. Sixty-eight patients had suffered a cardiac arrest, and 32 had had ventricular tachycardia or syncope. All patients except 9 underwent electrophysiological testing and had failed 1 +/- 1 drug trials before ICD implantation. Three types of nonthoracotomy lead systems were used. The nonthoracotomy lead with an ICD was successfully implanted in 96 patients (96%). Of the unsuccessful implants, 1 patient did not have venous access, the passive fixation lead in 1 would not remain lodged, 1 had elevated defibrillation thresholds, and 1 developed a hemopneumothorax while venous access was being obtained. The mean defibrillation threshold was 17 +/- 6 J. The mean procedure duration was 161 +/- 57 minutes. When a subcutaneous patch was used (n = 58), the procedure duration was 189 +/- 5 minutes, and when a subcutaneous patch was not required (n = 40), the procedure lasted 123 +/- 37 minutes (P < .0001). Patients remained in the hospital 4.5 +/- 4.1 days after implantation, with no procedure-related deaths. Acute complications occurred in 10 patients; 2 had lead dislodgments, 1 with previous abdominal surgery had his abdominal cavity entered (without other complications) while the ICD pocket was being made, 1 had postoperative heart failure, 1 developed a large hematoma when anticoagulation therapy was initiated, 3 required reintubation because of excessive anesthesia, 1 developed superficial cellulitis, and 1 developed a hemopneumothorax secondary to a lacerated subclavian vein. During 6 +/- 3 months of follow-up, 2 patients developed lead fractures. CONCLUSIONS (1) Electrophysiologists can implant an ICD with a nonthoracotomy lead system safely and with a high success rate; (2) use of a subcutaneous patch correlates with longer procedure durations; and (3) special precautions should be taken in patients with previous abdominal surgery.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 63 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Sedation for Electrophysiological Procedures;Pacing and Clinical Electrophysiology;2014-02-20

2. Evaluation of Defibrillation Efficacy at Implantation is Unnecessary;Cardiac Electrophysiology Clinics;2011-12

3. Permanent Pacemaker and Implantable Cardioverter-Defibrillator Implantation;Clinical Cardiac Pacing, Defibrillation and Resynchronization Therapy;2011

4. Implantation von Kardioverter-Defibrillatoren;Der Anaesthesist;2010-05-12

5. Intravenous sedation for cardiac procedures can be administered safely and cost-effectively by non-anesthesia personnel;Journal of Interventional Cardiac Electrophysiology;2008-01