Adverse Events With Transvenous Implantable Cardioverter-Defibrillators

Author:

Rosenqvist Mårten1,Beyer Thorsten1,Block Michael1,den Dulk Karel1,Minten Jaak1,Lindemans Fred1

Affiliation:

1. From Karolinska Hospital, Stockholm, Sweden (M.R.); Ruprecht-Karls University of Heidelberg, Heidelberg, Germany (T.B.); Westfälische Wilhelms Universität, Münster, Germany (M.B.); Academic Hospital Maastricht, the Netherlands (K.d.D.); and Bakken Research Center, Maastricht, the Netherlands (J.M., F.L.).

Abstract

Background —A newly developed classification system relates adverse events to the surgical procedure or the function of the implantable defibrillator. Methods and Results —Adverse events were monitored during prospective clinical evaluation of the Medtronic model 7219 Jewel ICD and were classified according to the definitions of the ISO 14155 standard for device clinical trials into 3 groups: severe and mild device-related and severe non–device-related adverse events. In addition, events were related to the surgical procedure, treatment with the device, or cardiac function. Seven hundred seventy-eight patients were followed up for an average of 4.0 months after ICD implantation. In total, 356 adverse events were observed in 259 patients. At 1, 3, and 12 months after ICD implantation, 99%, 98%, and 97% of the patients, respectively, survived; 95%, 93%, and 92%, respectively, were free of surgical reintervention; and 79%, 68%, and 51%, respectively, were free of any adverse event. Twenty patients died: 6 deaths were related to the surgical procedure, 12 deaths were considered unrelated to ICD treatment, and 2 patients died of an unknown cause. Of 111 nonlethal severe adverse device effects, 47 required surgical intervention, 19 times for correction of a dislodged lead. Inappropriate delivery of therapy was observed 128 times in 111 patients, and the events were typically resolved by reprogramming or drug adjustment. Nine of these required rehospitalization. Conclusions —Approximately 50% of patients experience an adverse event within the first year after ICD implantation. The observed adverse event rate depends on the definitions and the prospective monitoring. The incidence of inappropriate therapy emphasizes the need for improved detection algorithms and for quality-of-life evaluations, especially when considering ICD treatment in high-risk but arrhythmia-free patients.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Reference27 articles.

1. Complications of Implantable Cardioverter Defibrillator Therapy: Follow-Up of 241 Patients

2. EN540 European Standard. Clinical Investigation of Medical Devices for Human Subjects . Brussels Belgium: European Committee for Standardization; 1993.

Cited by 209 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献