Abstract
Background
Pulse wave distortions, mainly caused by reflections, and pressure gradients, caused by flow in the resistive vascular tree, may cause differences between finger and brachial artery pressures. These differences may limit the use of finger pressure measurements. We investigated whether brachial artery pressure waves could be reconstructed from finger pressure measurements by correcting for the pressure gradient in addition to correction for pulse wave distortion with a previously described filter.
Methods and Results
Finger artery pressure (with Finapres), intra-arterial brachial artery pressure (BAP), Riva-Rocci/Korotkoff (RRK), oscillometric, and return-to-flow (RTF) measurements were simultaneously performed in 57 healthy elderly subjects and patients with vascular disease and/or hypertension. A generalized waveform filter was used to correct for pulse wave distortions. Correction equations for the pressure gradient, based on finger pressure, RRK, RTF, or oscillometric measurements, were obtained in 28 randomly selected subjects and tested in 29. Before reconstruction, Finapres underestimated mean and diastolic BAP (finger pressure minus BAP: systolic, −3.2±16.9 mm Hg; mean, −13.0±10.5 mm Hg; diastolic, −8.4±9.0 mm Hg [mean±SD]). After filtering, reconstructed BAP waves were similar to actual BAP in shape but not in pressure level. Optimal correction for the pressure gradient with an equation based on RTF measurements reduced the pressure differences to meet American Association for the Advancement of Medical Instrumentation criteria (reconstructed finger pressure minus BAP: systolic, 3.7±7.0 mm Hg; mean, 0.7±4.6 mm Hg; and diastolic, 1.0±4.9 mm Hg).
Conclusions
BAP waves can be reconstructed from noninvasive finger pressure registrations when finger pressure waves are corrected for pulse wave distortion and individual pressure gradients.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Physiology (medical),Cardiology and Cardiovascular Medicine
Cited by
156 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献