1. From Harvard-MIT Division of Health Sciences and Technology, Massachusetts Institute of Technology, Cambridge (C.-W.H., D.W., E.R.E.), and the Cardiovascular Division, Department of Medicine, Brigham and Women’s Hospital, Harvard Medical School, Boston (E.R.E.), Mass.
Abstract
Background
The first compounds considered for stent-based delivery, such as heparin, were chosen on the basis of promising tissue culture and animal experiments, and yet they have failed to stop restenosis clinically. More recent compounds, such as paclitaxel, are of a different sort, being hydrophobic in nature, and their effects after local release seem far more profound. This dichotomy raises the question of whether drugs that have an effect when released from a stent do so because of differences in biology or differences in physicochemical properties and targeting.
Methods and Results
We applied continuum pharmacokinetics to examine the effects of transport forces and device geometry on the distribution of stent-delivered hydrophilic and hydrophobic drugs. We found that stent-based delivery invariably leads to large concentration gradients, with drug concentrations ranging from nil to several times the mean tissue concentration over a few micrometers. Concentration variations were a function of the Peclet number (Pe), the ratio of convective to diffusive forces. Although hydrophobic drugs exhibited greater variability than hydrophilic drugs, they achieved higher mean concentrations and remained closer to the intima. Inhomogeneous strut placement influenced hydrophilic drugs more negatively than hydrophobic drugs, dramatically affecting local concentrations without changing mean concentrations.
Conclusions
Because local concentrations and gradients are inextricably linked to biological effect, our results provide a potential explanation for the variable success of stent-based delivery. We conclude that mere proximity of delivery devices to tissues does not ensure adequate targeting, because physiological transport forces cause local concentrations to deviate significantly from mean concentrations.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Physiology (medical),Cardiology and Cardiovascular Medicine
Cited by
355 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献