Novel Passive Implantable Atrial Defibrillator Using Transcutaneous Radiofrequency Energy Transmission Successfully Cardioverts Atrial Fibrillation

Author:

Manoharan Ganesh1,Evans Noel1,Kidwai Bakthiar1,Allen Desmond1,Anderson John1,Adgey Jennifer1

Affiliation:

1. From the Regional Medical Cardiology Centre, Royal Victoria Hospital, Belfast (G.M., B.K., D.A., J. Adgey); the Department of Engineering, University of Ulster, Jordanstown, Northern Ireland (G.M., N.E., J. Anderson); and the Department of Physiology, Queen’s University of Belfast, Belfast (G.M., B.K., D.A.), UK.

Abstract

Background— Conventional methods for cardioversion of atrial fibrillation (AF) to sinus rhythm have numerous difficulties. A novel method for cardioversion using the passive implantable atrial defibrillator (PIAD) was tested in acute animal models. This device does not have a battery or a capacitor to store energy and is activated by transferring RF energy across the skin from an external transmitter to the subcutaneously implanted defibrillator. On activation, a novel monophasic shock waveform with 5% tilt is delivered to the heart via 2 intracardiac defibrillation leads. Methods and Results— Cardioversion attempts with the device were assessed in 2 phases: a feasibility and efficacy study and randomized comparison against standard waveforms. Defibrillation leads were placed transvenously into the distal coronary sinus and the right atrial appendage. These were connected to the subcutaneously implanted PIAD. Sustained AF was induced by rapid atrial pacing. The transmitter coil was placed on the skin overlying the defibrillator, and defibrillation synchronized to the R wave was attempted. The method was found to be efficacious at very low voltage and energy, with 100% cardioversion success observed for 10-ms 100-V shocks (mean energy, 1.54±0.02 J). The PIAD waveform had a higher cardioversion success rate than a truncated, 70% tilt monophasic exponential pulse (100 V, 100% versus 78.0±7.57%; P =0.001). There were no postshock complications. Conclusions— Considering these animal results, this method is promising for cardioverting AF in symptomatic patients.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3