Instantaneous Amplitude and Frequency Modulations Detect the Footprint of Rotational Activity and Reveal Stable Driver Regions as Targets for Persistent Atrial Fibrillation Ablation

Author:

Quintanilla Jorge G.123,Alfonso-Almazán José Manuel1,Pérez-Castellano Nicasio23,Pandit Sandeep V.4,Jalife José134,Pérez-Villacastín Julián123,Filgueiras-Rama David123

Affiliation:

1. From the Myocardial Pathophysiology Area, Centro Nacional de Investigaciones Cardiovasculares (CNIC), Madrid, Spain (J.G.Q., J.M.A.-A., J.J., D.F.-R.)

2. Arrhythmia Unit, Cardiology Department, Cardiovascular Institute, Instituto de Investigación Sanitaria del Hospital Clínico San Carlos (IdISSC), Madrid, Spain (J.G.Q., N.P.-C., J.P.-V., D.F.-R.)

3. Centro de Investigación Biomédica en Red de Enfermedades Cardiovasculares (CIBERCV), Madrid, Spain (J.G.Q., N.P.-C., J.J., J.P.-V., D.F.-R.)

4. Center for Arrhythmia Research, Department of Internal Medicine, University of Michigan, Ann Arbor (S.V.P., J.J.).

Abstract

Rationale: Costly proprietary panoramic multielectrode (64–256) acquisition systems are being increasingly used together with conventional electroanatomical mapping systems for persistent atrial fibrillation (PersAF) ablation. However, such approaches target alleged drivers (rotational/focal) regardless of their activation frequency dynamics. Objectives: To test the hypothesis that stable regions of higher than surrounding instantaneous frequency modulation (iFM) drive PersAF and determine whether rotational activity is specific for such regions. Methods and Results: First, novel single-signal algorithms based on instantaneous amplitude modulation (iAM) and iFM to detect rotational-footprints without panoramic multielectrode acquisition systems were tested in 125 optical movies from 5 ex vivo Langendorff-perfused PersAF sheep hearts (sensitivity/specificity, 92.6/97.5%; accuracy, 2.5-mm) and in computer simulations. Then, 16 pigs underwent high-rate atrial pacing to develop PersAF. After a median (interquartile range [IQR]) of 4.4 (IQR, 2.5–9.9) months of high-rate atrial pacing followed by 4.1 (IQR, 2.7–5.4) months of self-sustained PersAF, pigs underwent in vivo high-density electroanatomical atrial mapping (4920 [IQR, 4435–5855] 8-second unipolar signals per map). The first 4 out of 16 pigs were used to adapt ex vivo optical proccessing of iFM/iAM to in vivo electrical signals. In the remaining 12 out of 16 pigs, regions of higher than surrounding average iFM were considered leading-drivers. Two leading-driver + rotational-footprint maps were generated 2.6 (IQR, 2.4–2.9) hours apart to test leading-driver spatiotemporal stability and guide ablation. Leading-driver regions (2.5 [IQR, 2.0–4.0] regions/map) exactly colocalized (95.7%) in the 2 maps, and their ablation terminated PersAF in 92.3% of procedures (radiofrequency until termination, 16.9 [IQR, 9.2–35.8] minutes; until nonsustainability, 20.4 [IQR, 12.8–44.0] minutes). Rotational-footprints were found at every leading-driver region, albeit most (76.8% [IQR, 70.5%–83.6%]) were located outside. Finally, the translational ability of this approach was tested in 3 PersAF redo patients. Conclusions: Both rotational-footprints and spatiotemporally stable leading-driver regions can be located using iFM/iAM algorithms without panoramic multielectrode acquisition systems. In pigs, ablation of leading-driver regions usually terminates PersAF and prevents its sustainability. Rotational activations are sensitive but not specific to such regions. Single-signal iFM/iAM algorithms could be integrated into conventional electroanatomical mapping systems to improve driver detection accuracy and reduce the cost of patient-tailored/mechanistic approaches.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Cardiology and Cardiovascular Medicine,Physiology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3