Abstract
Dynamic characteristics of a variety of catheters and their dependence on various operating parameters are presented. Careful flushing of the catheter with degassed water or saline considerably improves the performance of these systems. The operating temperature, number of catheter uses, and average transmural pressure difference do not have a significant effect.
The error due to catheter distortion in commonly used left ventricular performance indices (end diastolic pressure, peak systolic pressure, maximum dp/dt and V
max
) was assessed. On the basis of these results, it is shown that the natural frequency of the catheter system must be greater than 40 Hz to produce results accurate to 10%.
The experimental data were used to develop a semi-empirical model for the catheter transducer system. This model was used in the design of an analog compensator to further improve the dynamic characteristics of the catheter transducer system. The compensator requires only two parameter settings, catheter natural frequency and damping ratio.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Physiology (medical),Cardiology and Cardiovascular Medicine
Reference15 articles.
1. Blood pressure measuring methods;IRE Trans Biomed Elect,1972
2. A hydraulic pressure generator for testing the dynamic characteristics of blood pressure manometers;J Lab Clin Med,1959
3. Physiologic Recording by Modern Instruments With Particular Reference to Pressure Recording
4. A Critical Study of the Response of Manometers to Forced Oscillations
Cited by
57 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献