Comparison of Electroanatomic Contact and Noncontact Mapping of Ventricular Scar in a Postinfarct Ovine Model With Intramural Needle Electrode Recording and Histological Validation

Author:

Sivagangabalan Gopal1,Pouliopoulos Jim1,Huang Kaimin1,Lu Juntang1,Barry Michael A.1,Thiagalingam Aravinda1,Ross David L.1,Thomas Stuart P.1,Kovoor Pramesh1

Affiliation:

1. From the Cardiology Department, Westmead Hospital, Sydney, Australia.

Abstract

Background— Substrate-based ablation is useful for nonhemodynamically tolerated postinfarct ventricular tachycardia. We assessed the accuracy of the CARTO contact and EnSite noncontact systems at identifying scar in a chronic ovine model with intramural plunge needle electrode recording and histological validation. Methods and Results— Scar mapping was performed on 8 male sheep with previous percutaneous-induced myocardial infarction. Up to 20 plunge needles were inserted into the left ventricle of each animal in areas of dense scar, scar border, and normal myocardium. A simultaneous CARTO map and EnSite geometry were acquired using a single catheter, and needle electrode locations were registered. A dynamic substrate map was constructed using ratiometric 50% peak negative voltage. The scar percentage around each needle location was quantified histologically. Analysis was performed on 152 plunge needles and corresponding histological blocks. Spearman correlation with histology was 0.690 ( P <0.001) for needle electrode peak-to-peak voltage (PPV), 0.362 ( P <0.001) and 0.492 ( P <0.001) for CARTO bipolar and unipolar PPV, and 0.381 ( P <0.001) for EnSite dynamic substrate map (≤40 mm from array). The area under the receiver operator characteristics curve (<50% and ≥50% scar) was 0.896 for needle electrode PPV, 0.726 and 0.697 for CARTO bipolar and unipolar PPV, and 0.703 for EnSite dynamic substrate map (≤40 mm from array). Conclusions— Both the CARTO contact and EnSite noncontact systems were moderately accurate in identifying postinfarct scar when compared with intramural electrodes and confirmed with histology. The EnSite dynamic substrate map was comparable to the CARTO contact bipolar PPV when points >40 mm from the array were excluded.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 20 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3