Affiliation:
1. From the Department of Cardiovascular Electrophysiology, Heart and Vascular Institute, University of Pittsburgh Medical Center, PA.
Abstract
Background—
Current implantable cardiac devices have a finite battery life of ≈3 to 7 years for implantable cardioverter-defibrillators. It is current practice to reuse all properly functioning intravascular leads. We tested the hypothesis that a strategy of prophylactic lead removal at the time of device change would be superior under some conditions to the current practice of lead reuse.
Methods and Results—
Using currently available data and a Monte Carlo microsimulation trial, we calculated the risks of leaving an indwelling lead until extraction is indicated because of malfunction versus an aggressive management strategy of prophylactic serial extraction at time of generator change. With a serial lead exchange strategy of leads at generator change, there is reduced overall extraction-related mortality because of fewer late complications attributable to extraction of leads with high dwell time because of infection, recall, or subsequent lead failure. This finding is limited to young patients or those with high expected indwell time of lead. This trend reverses for leads with <40 years expected dwell time. Sensitivity analysis shows high dependence on extraction performance and device longevity. In all cases, serial extraction would be expected to lead to increased adverse events related to the more complex procedure.
Conclusions—
A strategy of serial lead extraction, given best available current parameters, yields a lower procedural mortality risk in the long-term management of indwelling implantable cardioverter-defibrillator leads in young patients (>40-year estimated dwell time) driven by high aggregate anticipated risk of lifetime lead complication.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Physiology (medical),Cardiology and Cardiovascular Medicine
Cited by
11 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献