Characterization of Phrenic Nerve Response to Pulsed Field Ablation

Author:

Howard Brian1,Haines David E.2,Verma Atul3ORCID,Kirchhof Nicole1ORCID,Barka Noah1ORCID,Onal Birce1ORCID,Stewart Mark T.1ORCID,Sigg Daniel C.1ORCID

Affiliation:

1. Medtronic, Minneapolis, MN (B.H., N.K., N.B., B.O., M.T.S., D.C.S.).

2. Beaumont, Royal Oaks, MI (D.E.H.).

3. McGill University Health Center, McGill University, Montreal, Quebec, Canada (A.V.).

Abstract

Background: Phrenic nerve palsy is a well-known complication of cardiac ablation, resulting from the application of direct thermal energy. Emerging pulsed field ablation (PFA) may reduce the risk of phrenic nerve injury but has not been well characterized. Methods: Accelerometers and continuous pacing were used during PFA deliveries in a porcine model. Acute dose response was established in a first experimental phase with ascending PFA intensity delivered to the phrenic nerve (n=12). In a second phase, nerves were targeted with a single ablation level to observe the effect of repetitive ablations on nerve function (n=4). A third chronic phase characterized assessed histopathology of nerves adjacent to ablated cardiac tissue (n=6). Results: Acutely, we observed a dose-dependent response in phrenic nerve function including reversible stunning (R 2 =0.965, P <0.001). Furthermore, acute results demonstrated that phrenic nerve function responded to varying levels of PFA and catheter proximity placements, resulting in either: no effect, effect, or stunning. In the chronic study phase, successful isolation of superior vena cava at a dose not predicted to cause phrenic nerve dysfunction was associated with normal phrenic nerve function and normal phrenic nerve histopathology at 4 weeks. Conclusions: Proximity of the catheter to the phrenic nerve and the PFA dose level were critical for phrenic nerve response. Gross and histopathologic evaluation of phrenic nerves and diaphragms at a chronic time point yielded no injury. These results provide a basis for understanding the susceptibility and recovery of phrenic nerves in response to PFA and a need for appropriate caution in moving beyond animal models.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3