Feasibility of Real-Time MRI With a Novel Carbon Catheter for Interventional Electrophysiology

Author:

Nordbeck Peter1,Bauer Wolfgang R.1,Fidler Florian1,Warmuth Marcus1,Hiller Karl-Heinz1,Nahrendorf Matthias1,Maxfield Michelle1,Wurtz Sabine1,Geistert Wolfgang1,Broscheit Jens1,Jakob Peter M.1,Ritter Oliver1

Affiliation:

1. From the Department of Internal Medicine I (P.N., W.R.B., O.R.), the Department of Experimental Physics V (P.N., M.W., P.M.J.), and the Department of Anesthesiology (J.B.), University of Würzburg, Germany; Research Center Magnetic-Resonance-Bavaria (F.F., K.-H.H.), Würzburg, Germany; Center for Molecular Imaging Research (M.N.), Massachusetts General Hospital, Harvard Medical School, Boston, Mass; and Biotronik GmbH & Co. KG (M.M., S.W., W.G.), Berlin, Germany.

Abstract

Background— Cardiac MRI offers 3D real-time imaging with unsurpassed soft tissue contrast without x-ray exposure. To minimize safety concerns and imaging artifacts in MR-guided interventional electrophysiology (EP), we aimed at developing a setup including catheters for ablation therapy based on carbon technology. Methods and Results— The setup, including a steerable carbon catheter, was tested for safety, image distortion, and feasibility of diagnostic EP studies and radiofrequency ablation at 1.5 T. MRI was performed in 3 different 1.5-T whole-body scanners using various receive coils and pulse sequences. To assess unintentional heating of the catheters by radiofrequency pulses of the MR scanner in vitro, a fluoroptic thermometry system was used to record heating at the catheter tip. Programmed stimulation and ablation therapy was performed in 8 pigs. There was no significant heating of the carbon catheters while using short, repetitive radiofrequency pulses from the MR system. Because there was no image distortion when using the carbon catheters, exact targeting of the lesion sites was possible. Both atrial and ventricular radiofrequency ablation procedures including atrioventricular node modulation were performed successfully in the scanner. Potential complications such as pericardial effusion after intentional perforation of the right ventricular free wall during ablation could be monitored in real time as well. Conclusion— We describe a newly developed EP technology for interventional electrophysiology based on carbon catheters. The feasibility of this approach was demonstrated by safety testing and performing EP studies and ablation therapy with carbon catheters in the MRI environment.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 34 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Real-time cardiovascular magnetic resonance-guided radiofrequency ablation: A comprehensive review;World Journal of Cardiology;2023-09-26

2. Cardiovascular Magnetic Resonance-Guided Radiofrequency Ablation;JACC: Clinical Electrophysiology;2022-02

3. Advances in Real-Time MRI–Guided Electrophysiology;Current Cardiovascular Imaging Reports;2019-02

4. Real-time magnetic resonance imaging-guided cardiac electrophysiology: the long road to clinical routine;European Heart Journal - Cardiovascular Imaging;2018-11-30

5. MR-guided Cardiac Interventions;Topics in Magnetic Resonance Imaging;2018-06