Reimplantation and Repeat Infection After Cardiac-Implantable Electronic Device Infections

Author:

Boyle Thomas A.1,Uslan Daniel Z.1,Prutkin Jordan M.1,Greenspon Arnold J.1,Baddour Larry M.1,Danik Stephan B.1,Tolosana Jose M.1,Le Katherine1,Miro Jose M.1,Peacock James1,Sohail Muhammad R.1,Vikram Holenarasipur R.1,Carrillo Roger G.1

Affiliation:

1. From the Department of Surgery, Division of Cardiothoracic Surgery, University of Miami, FL (T.A.B.); Department of Medicine, Division of Infectious Diseases, David Geffen School of Medicine at University of California, Los Angeles (D.Z.U.); Department of Medicine, Division of Cardiology, University of Washington, Seattle (J.M.P.); Department of Medicine, Division of Cardiology, Cardiac Electrophysiology Laboratory, Thomas Jefferson University Hospital, Philadelphia, PA (A.J.G.); Department of...

Abstract

Background— Infection is a serious complication of cardiovascular-implantable electronic device implantation and necessitates removal of all hardware for optimal treatment. Strategies for reimplanting hardware after infection vary widely and have not previously been analyzed using a large, multicenter study. Methods and Results— The MEDIC (Multicenter Electrophysiologic Device Infection Cohort) prospectively enrolled subjects with cardiovascular-implantable electronic device infections at multiple institutions in the United States and abroad between 2009 and 2012. Reimplantation strategies were evaluated overall, and every patient who relapsed within 6 months was individually examined for clinical information that could help explain the negative outcome. Overall, 434 patients with cardiovascular-implantable electronic device infections were prospectively enrolled at participating centers. During the initial course of therapy, complete device removal was done in 381 patients (87.8%), and 220 of them (57.7%) were ultimately reimplanted with new devices. Overall, the median time between removal and reimplantation was 10 days, with an interquartile range of 6 to 19 days. Eleven of the 434 patients had another infection within 6 months, but only 4 of them were managed with cardiovascular-implantable electronic device removal and reimplantation during the initial infection. Thus, the repeat infection rate was low (1.8%) in those who were reimplanted. Patients who retained original hardware had a 11.3% repeat infection rate. Conclusions— Our study findings confirm that a broad range of reimplant strategies are used in clinical practice. They suggest that it is safe to reimplant cardiac devices after extraction of previously infected hardware and that the risk of a second infection is low, regardless of reimplant timing.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370