A Novel Nonobstructive Intravascular MRI Coil

Author:

Worthley Stephen G.1,Helft Gérard1,Fuster Valentin1,Fayad Zahi A.1,Shinnar Meir1,Minkoff Lawrence A.1,Schechter Clyde1,Fallon John T.1,Badimon Juan J.1

Affiliation:

1. From the Zena and Michael A. Wiener Cardiovascular Institute (S.G.W., G.H., V.F., Z.A.F., M.S., J.T.F., J.J.B.), Mount Sinai School of Medicine; Magna-Lab, Inc (L.A.M.); and Department of Family Medicine and Community Health (C.S.), Albert Einstein College of Medicine, New York, NY.

Abstract

Objective— MRI is being used to characterize the composition of atherosclerotic plaques. However, the resolution achievable using surface radiofrequency coils is limited by the signal to noise ratio. We studied the utility of a new intravascular (IV) MRI probe for high-resolution in vivo imaging of atherosclerotic lesions. Methods and Results— Balloon-injured Watanabe heritable hyperlipidemic rabbits served as the experimental model of atherosclerosis. The newly developed IV MRI probe is 1.3 mm in diameter and can be positioned over a guidewire. MRI was performed with both an external phased-array coil and the IV MR coil. MR observations were correlated with histopathology. After MRI, the animals were killed and analysis of agreement between MR and histopathology was performed. The IV MR coil allows aortic images to be obtained with 156×156 μm 2 in-plane resolution versus 352×352 μm 2 when used with the external phased-array coil. No significant motion artifacts were noted, despite the continuation of arterial blood flow during image acquisition around the IV probe. The different components of the atherosclerotic lesions (lipid core and fibrous cap) were easily identified. There was an excellent agreement between MRI with the IV coil and histopathology by simple linear regression for both the mean wall thickness ( r =0.88, slope 0.82, P <0.0001) and vessel wall area ( r =0.86, slope 1.08, P <0.0001). Conclusions— The new nonobstructive design for this intravascular coil provides great promise for additional work in high-resolution MRI characterization of atherosclerotic plaques in vivo. The ability to position the probe with a guidewire allows its placement under fluoroscopic or MRI guidance, whereas its size is compatible with human coronary arteries.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Cardiology and Cardiovascular Medicine

Cited by 41 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3