Inaccuracy of mitral pressure half-time immediately after percutaneous mitral valvotomy. Dependence on transmitral gradient and left atrial and ventricular compliance.

Author:

Thomas J D1,Wilkins G T1,Choong C Y1,Abascal V M1,Palacios I F1,Block P C1,Weyman A E1

Affiliation:

1. Noninvasive Cardiac Laboratory, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts 02114.

Abstract

Mitral pressure half-time (T1/2) is widely used as an independent measure of mitral valve area in patients undergoing percutaneous mitral valvotomy. However, fluid dynamics theory predicts T1/2 to be strongly dependent on chamber compliance and the peak transmitral gradient, which are variables that change dramatically with valvotomy. These theoretical predictions were tested in an in vitro model of the left heart where valve area, chamber compliance, and initial gradient were independently adjusted. Measured T1/2 was observed to vary inversely with orifice area and directly with net chamber compliance and the square root of the initial pressure gradient. Theoretical predictions of T1/2 agreed with observed values with r = 0.998. To test this theory in vivo, the hemodynamic tracings of 18 patients undergoing mitral valvotomy were reviewed. Predictions were made for T1/2 assuming dependence only on valve area; these showed some correlations before valvotomy (r = 0.48-0.64, p less than 0.05) but none after valvotomy (r = 0.05-0.28, p = NS). Predictions for T1/2 based on the theoretical derivation (and thus including compliance and pressure in their calculation) were much better: before valvotomy, r = 0.93-0.96, p less than 0.0001; after valvotomy, r = 0.52-0.66, p less than 0.05. These data indicate that T1/2 is not an independent inverse measure of mitral valve area but is also directly proportional to net chamber compliance and the square root of the initial transmitral gradient. These other factors render T1/2 an unreliable measure of mitral valve area in the setting of acute mitral valvotomy.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3