Precision of measurements of right and left ventricular volume by cine computed tomography.

Author:

Reiter S J,Rumberger J A,Feiring A J,Stanford W,Marcus M L

Abstract

Precise determination of left and right ventricular stroke volumes is limited with conventional imaging techniques. We determined whether right and left ventricular stroke volumes could be precisely measured with cine computed tomography (CT). Cine CT enables simultaneous imaging of the right and left ventricles at an 8 mm slice thickness with a maximal scanning rate of 17 frames/sec (50 msec acquisition intervals). In eight dogs, true right ventricular and left ventricular stroke volumes were determined by dividing thermodilution cardiac output by heart rate and/or with the use of an aortic electromagnetic flow probe implanted over a long term. After at least 5 sec of suspended respiration, cine CT images were acquired during central venous injection of a nonionic contrast agent. Multiple perturbations in stroke volume were induced in each dog by the administration of dobutamine, sodium pentobarbital, or sodium nitroprusside or by coronary artery occlusion. Right and left ventricular stroke volumes were obtained by Simpson's reconstruction of end-diastolic and end-systolic short-axis tomograms from apex to base. The cine CT left ventricular stroke volume (range 11 to 45 ml) correlated highly with the true left ventricular stroke volume (r = .99, slope = 1.01, y intercept = -0.2 ml, SEE = 1.5 ml, n = 25). The cine CT right ventricular stroke volume (range 11 to 34 ml) also correlated highly with the true right ventricular stroke volume (r = .98, slope = 0.9, y intercept = 2.2 ml, SEE = 1.7 ml, n = 15). In 12 studies, the mean difference between nearly simultaneous right and left ventricular stroke volumes by cine CT was 1.1 ml (range 0.1 to 3.2 ml). Calculation of right and left ventricular stroke volumes from data from cine CT were highly reproducible. Intraobserver variability in measurements of right ventricular stroke volume (r = 1.0, slope = 0.99, y intercept = 0.19 ml) and left ventricular stroke volume (r = 1.0, slope = 1.02, y intercept = -0.21 ml) was minimal. Interobserver variability in measurements of right ventricular stroke volume (r = .98, slope = 0.90, y intercept = 1.66 ml) and left ventricular stroke volume (r = .99, slope = 0.97, y intercept = -0.02 ml) was likewise minimal. Thus, precise and highly reproducible measurements of right and left ventricular stroke volumes can be obtained with cine CT.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 177 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. X‐ray Tomography and Tomoscopy on Metals: A Review;Advanced Engineering Materials;2023-01-10

2. Ventricular Enlargement;Clinical Electrocardiography;2021-09-17

3. Indikationsstellung zur Digitalen Volumentomographie;Digitale Volumentomographie;2021

4. Computed Tomograph Cardiovascular Imaging;Coronary Artery Disease;2015

5. Ventricular Enlargement;Clinical Electrocardiography;2012-03-22

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3