Abstract
The indirect oscillometric method of blood pressure measurement was compared to the direct arterial puncture method in 15 anesthetized dogs, divided into three weight groups, undergoing a variety of surgical procedures. The objectives of this study were to determine the accuracy of the indirect oscillometric method at a single point in time and when sequential values were averaged. Additionally, the ability to detect systemic hypotension (i.e., mean systemic arterial pressure less than 60 mmHg) was evaluated. The method had the highest correlation coefficient (r of 0.8) when five sequential values were averaged and compared, and it appeared to be sensitive (100%) and specific (91%) for detecting hypotension.