Verifying Placement of Small-Bore Feeding Tubes: Electromagnetic Device Images Versus Abdominal Radiographs

Author:

Bryant Vera1,Phang Jean1,Abrams Kevin1

Affiliation:

1. Vera Bryant is an acute care nurse practitioner in the neuroscience department at Baptist Hospital of Miami, Miami, Florida. Jean Phang is a quality assurance nurse at Baptist Hospital of Miami. Kevin Abrams is medical director of the radiology department at Baptist Hospital of Miami.

Abstract

Background Clinicians are unsure if radiography is needed to confirm correct positioning of feeding tubes inserted with assistance from an electromagnetic system. Objectives To compare radiographic reports of feeding tube placement with images generated by an electromagnetic feeding tube placement device. Methods The medical records of 200 consecutive patients who had feeding tubes inserted with assistance from an electromagnetic feeding tube placement device were reviewed retrospectively. Radiographic reports of tube site were compared with images generated by the device. Results Radiographic evidence of tube sites was available in 188 cases: 184 tubes were located in portions of the gastrointestinal tract. Ninety of the 188 tubes were situated in the optimal site (distal duodenum or jejunum) radiographically. Images generated by the electromagnetic device were available in 176 cases; of these, 52 tubes appeared to end in the expected left lower quadrant. Tubes shown on radiographs to be in other sites also occasionally appeared to end in the left lower quadrant. Nurses using the device did not recognize 4 of the 188 tubes (2.1%) that were inadvertently placed in the lung. No consistent pattern of quadrant distribution was found for tubes positioned in the stomach or proximal duodenum. Conclusions Images generated by the electromagnetic tube placement device provided inconsistent results regarding tube location. A small percentage of seriously malpositioned tubes were not detected by using the electromagnetic device. These findings do not support eliminating radiographs to confirm correct tube placement following use of an electromagnetic tube placement device.

Publisher

AACN Publishing

Subject

Critical Care,General Medicine

Cited by 29 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3