Abstract
AbstractQuantitative de terminations of thickness and composition are presented for 100- to 10,000-A iron-nickel films. Calibration of X-ray intensities with interferometric and calorimetric values give linear plots for films deposited on Mylar or glass substrates. Application of a film-casting technique shows promise of alleviating two of the problems encountered in X-ray spectroscopy, i.e., matrix-effect-caused deviations from proportionality and the lack of standards for calibration of many types of material.
Publisher
Cambridge University Press (CUP)
Reference9 articles.
1. Chemical Analysis of Thin Films by X-Ray Emission Spectrography
2. Trace Analysis by X-Ray Emission Spectrography
3. X-Ray Fluorescent Intensity of Elements Evaporated from Solution onto Thin Film
4. Month A. , Schindler M. , and Antales J. , “Measurement of Composition and Thickness of Iron-Nickel Films by X-Ray Fluorescence Techniques,” Pittsburgh Conference of Analytical Chemistry and Applied Spectroscopy, February-March 1961.