Affiliation:
1. Interacoustics A/S, Audiometer Allé 1, Middelfart, DK-5500, Denmark
Abstract
Measurements of wideband acoustic immittance (WAI) rely on the calibration of an ear probe to obtain its acoustic source parameters. The clinical use of WAI and instruments offering the functionality are steadily growing, however, no international standard exists to ensure a certain reliability of the hardware and methods underlying such measurements. This paper describes a reciprocity method that can evaluate the accuracy of and identify errors in ear-probe source calibrations. By placing the ear probes of two calibrated WAI instruments face-to-face at opposite ends of a short waveguide, the source parameters of each ear probe can be measured using the opposite calibrated ear probe. The calibrated and measured source parameters of each ear probe can then be compared directly, and the influence of possible calibration errors on WAI measurements may be approximated. In various exemplary ear-probe calibrations presented here, the reciprocity method accurately identifies errors that would otherwise remain undetected and result in measurement errors in real ears. The method is likely unsuitable for routine calibration of WAI instruments but may be considered for conformance testing as part of a potential future WAI standard.
Publisher
Acoustical Society of America (ASA)
Subject
Acoustics and Ultrasonics,Arts and Humanities (miscellaneous)
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献