Comparison between the Parks Medical Doppler and the Mano Médical Vet BP Doppler for recording indirect systemic blood pressure in conscious dogs

Author:

Domingues Marina,Archontakis Panagiotis,Bouvard Jonathan,Winter Luke,Moore Danielle

Abstract

Objectives: To evaluate agreement between Parks Medical Doppler (PMD) and Mano Médical Vet BP Doppler (MMVBP) for assessing systemic blood pressure (SBP) in conscious dogs. Methods: 40 client-owned dogs were prospectively enrolled; SBP measurements were acquired by a single operator using the PMD then the MMVBP. The mean of 5 consecutive measurements for each device was classified according to target organ damage (TOD) risk scores (1 < 140 mmHg; 2 =140-160 mmHg; 3 =160-180 mmHg; 4 ≥180 mmHg). Results: Total mean SBP for the devices was not statistically different (p =0.77). However, Bland-Altman analysis revealed wide limits of agreement, with MMVBP slightly underestimating SBP compared to PMD (bias = –0.6mmHg, 95% LoA: – 26.3 – 25.09). Both devices correlated well (r = 0.8269; p < 0.0001) and had identically acceptable intra-observer repeatability (coefficients of variation of 4.09% for MMVBP; 3.86% for PMD). Four dogs (10%) had a TOD score of 3 by one device but scored < 3 with the other. Conclusions: A good agreement and correlation was observed between the PMD and the MMVBP suggesting that both devices can be used interchangeably for assessment of SBP in conscious dogs. The wide limits of agreement observed between both devices were most likely associated with intraindividual variability in SBP over time.

Publisher

ScopeMed

Subject

General Veterinary

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3