A modified, optimized kinetic photometric thrombin generation assay components to measure the potency of thrombin in plasma

Author:

Zrari Salar A.

Abstract

The purpose of the present study was to improve, optimize and modification, a kinetic photometric assay of Factor IIa activity with regard to blood sample types, calibration, analytic corrections, and activation reagents, in order to establish a rapid routine, less time consuming, and suitable method for the evaluation of large numbers of clinical samples. Our modified method was based on the continuous monitoring of the Factor IIa released from the reacting substance, in the presence of a specific chromogenic peptide substrate (S2238). The optimization of pH and of the concentrations of substrate, activator (CaCl2), EDTA, and tissue factor was carried out as well as blood sample types (Platelet-rich and platelet-poor plasma samples) so as to ensure the best compromise between reliability, specificity, and sensitivity, However, the Factor IIa activity was compared in the plasma of different donors. In conclusion, our modified and optimized FIIa generation assay is a rapid and simple method for the diagnosis and monitoring of different clinical cases, at the same time it can be easily adapted to clinical chemistry analyzers.

Publisher

CMB Association

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3