Affiliation:
1. Department of Neurosurgery, Junshin Hospital, Kakogawa, Japan.
2. Department of Neurosurgery, Kobe University Graduate School of Medicine, Kobe, Japan.
Abstract
Background:
In the treatment of giant cerebral aneurysms with flow-diverting stents, access to the distal parent artery is critical but occasionally challenging. This article provides our experience with a novel steerable microcatheter in such a situation, as well as a review of the literature.
Case Description:
A 73-year-old woman presented with right ptosis and external ophthalmoplegia. Magnetic resonance angiography revealed a giant right cavernous internal carotid artery aneurysm. Endovascular treatment was planned with flow diversion, but distal access was not possible using the standard technique. A 2.4-Fr steerable microcatheter, Leonis Mova Selective, was implemented, and by bending the catheter tip toward the distal parent artery, a guidewire could be guided distally. After the catheter exchange, two flow-diverting stents were deployed successfully.
Conclusion:
Steerable microcatheters may provide an option in treatment with flow-diverting stents for giant cerebral aneurysms where access to the distal parent artery is compromised.