Objectives:
With the increasing threat of multidrug-resistant organisms, colistin has become popular in clinical practice. A better understanding of antimicrobial susceptibility testing methods for colistin is needed for optimal patient management. The aim of the study was to determine the accuracy of E-test, Vitek 2 system for the detection of colistin minimum inhibitory concentrations (MIC) against broth microdilution (BMD).
Material and Methods:
A total of 100 isolates of Gram-negative bacilli were subjected to susceptibility testing for colistin using the following methods: BMD, E-test, Vitek 2, and disk diffusion. Using BMD as the gold standard, comparative analysis between different methods was carried out.
Results:
Comparison of MIC values of E-test (GM = 0.488 mg/ml) against BMD (GM = 0.611 mg/ml using unpaired t-test (t = 2.015, P = 0.045) showed that geometric means of MIC values of E-strip were significantly lower than BMD. Similarly, comparison of MIC values of Vitek 2 system (GM = 0.615 mg/ml) against BMD (GM = 0.611 mg/ml) using unpaired t-test (t = −0.050, P = 0.960) showed no statistical significant differences in geometric means of MIC values. Taking reference as BMD method – the EA for E-strip is 57%, CA is 97%, VME is 2%, and no ME. Similarly, for the Vitek method EA is 64%, CA is 98%, VME is 1%, and ME is 1%.
Conclusion:
Different susceptibility testing methods for colistin show great variation in their results and BMD is the best candidate as gold standard. The Vitek 2 method showed good concordance with BMD.