Affiliation:
1. Department of Cardiac Anaesthesia and Critical Care, All India Institute of Medical Sciences, New Delhi, India,
Abstract
Objective:
To identify the correlation between systolic time ratio(STR) measured with electrical cardiometry (EC) device ICON (Osypka Medical, Berlin, Germany) and echocardiography in patients undergoing cardiac surgery and to find a noninvasive technique for estimating left ventricular systolic function during the perioperative period.
Materials and Methods:
Systolic time ratio data were obtained simultaneously from the electrical cardiometry device ICON (Osypka Medical, Berlin, Germany) and transthoracic echocardiography at the following predefined timepoints— before anaesthesia induction (T0 baseline), after induction (T1), at the end of the surgery(T2), and after extubation (T3) in patients undergoing elective cardiac surgery. The agreement between the systolic time ratio measured by electrical cardiometry and transthoracic echocardiography was studied using Bland-Altman plots. Paired t-tests were used to compare systolic time ratio measurements at different time points.
Results:
Mean STR by EC and Echocardiography at T0 was 0.456 (0.429-0.483) and 0.348 (0.330-.366) at T1 was 0.464 (0.442 -0.486) and 0.372 (0.344-0.401) at T2 was 0.421 (0.402 -0.439) and 0.305 (0.290-0.320) and at T3 was 0.438 (0.419-0.457) and 0.353 (0.336-0.370), P value <0.001. Bland-Altman analysis showed that EC measured STR compared with echocardiography at T0 with a mean bias of 0.108 and (with limits of agreement -0.19 ,0.14) at T1 it was 0.092 (with limits of agreement -0.21,0.40) at T2 it was 0.11 (with limits of agreement -0.04,0.28) and at T3 it was 0.085 (with limits of agreement -0.101 ,0.271).
Conclusion:
In conclusion, no association between systolic time ratio as measured by electrical cardiometry and echocardiography was found in our study. Our results do not conclusively prohibit using electrical cardiometry in the perioperative period to evaluate LV systolic function. Further work must be done to establish the role of STR as a surrogate marker of LV systolic function.