Affiliation:
1. Servicio de Microbiologı́a, Hospital Ramón y Cajal, Madrid 28034, Spain
Abstract
ABSTRACT
The Wider system is a newly developed computer-assisted image-processing device for both bacterial identification and antimicrobial susceptibility testing. It has been adapted to be able to read and interpret commercial MicroScan panels. Two hundred forty-four fresh consecutive clinical isolates (138 isolates of the family
Enterobacteriaceae
, 25 nonfermentative gram-negative rods [NFGNRs], and 81 gram-positive cocci) were tested. In addition, 100 enterobacterial strains with known β-lactam resistance mechanisms (22 strains with chromosomal AmpC β-lactamase, 8 strains with chromosomal class A β-lactamase, 21 broad-spectrum and IRT β-lactamase-producing strains, 41 extended-spectrum β-lactamase-producing strains, and 8 permeability mutants) were tested. API galleries and National Committee for Clinical Laboratory Standards (NCCLS) microdilution methods were used as reference methods. The Wider system correctly identified 97.5% of the clinical isolates at the species level. Overall essential agreement (±1 log
2
dilution for 3,719 organism-antimicrobial drug combinations) was 95.6% (isolates of the family
Enterobacteriaceae
, 96.6%; NFGNRs, 88.0%; gram-positive cocci, 95.6%). The lowest essential agreement was observed with
Enterobacteriaceae
versus imipenem (84.0%), NFGNR versus piperacillin (88.0%) and cefepime (88.0%), and gram-positive isolates versus penicillin (80.4%). The category error rate (NCCLS criteria) was 4.2% (2.0% very major errors, 0.6% major errors, and 1.5% minor errors). Essential agreement and interpretive error rates for eight β-lactam antibiotics against isolates of the family
Enterobacteriaceae
with known β-lactam resistance mechanisms were 94.8 and 5.4%, respectively. Interestingly, the very major error rate was only 0.8%. Minor errors (3.6%) were mainly observed with amoxicillin-clavulanate and cefepime against extended-spectrum β-lactamase-producing isolates. The Wider system is a new reliable tool which applies the image-processing technology to the reading of commercial trays for both bacterial identification and susceptibility testing.
Publisher
American Society for Microbiology
Reference31 articles.
1. Evaluation of the MicroScan antimicrobial susceptibility system with the AutoScan-4 automated reader;Baker C. N.;J. Clin. Microbiol.,1986
2. Criterios del Grupo MENSURA para la definición de los puntos crı́ticos de sensibilidad a los antibióticos;Baquero F.;Rev. Esp. Quimioter.,1997
3. Comparison of efficacy and cost-effectiveness of BIOMIC VIDEO and Vitek antimicrobial susceptibility test systems for use in the clinical microbiology laboratory
4. A critique of semiautomated susceptibility systems;Cherubin C.;Rev. Infect. Dis.,1987
5. Imipenem stability in a predried susceptibility panel
Cited by
45 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献