Author:
Robertson J H,Gleason D,Tsuji K
Abstract
A dry-heat oven with automatic, multiple-sample introduction and withdrawal has been constructed to achieve instantaneous heating and cooling of samples. The oven temperature fluctuation at set points of 170 to 250 degrees C was +/- 0.1 degrees C, with temperature variation between the replicate samples of +/- 0.2 degrees C. Correction required for a sample come-up time was minimal, i.e., less than 0.25 min of the dry-heat destruction time.
Publisher
American Society for Microbiology
Subject
Ecology,Applied Microbiology and Biotechnology,Food Science,Biotechnology
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献