Analysis of the disagreement between automated bioluminescence-based and culture methods for detecting significant bacteriuria, with proposals for standardizing evaluations of bacteriuria detection methods

Author:

Nichols W W,Curtis G D,Johnston H H

Abstract

A fully automated method for detecting significant bacteriuria is described which uses firefly luciferin and luciferase to detect bacterial ATP in urine. The automated method was calibrated and evaluated, using 308 urine specimens, against two reference culture methods. We obtained a specificity of 0.79 and sensitivity of 0.75 using a quantitative pour plate reference test and a specificity of 0.79 and a sensitivity of 0.90 using a semiquantitative standard loop reference test. The majority of specimens negative by the automated test but positive by the pour plate reference test were specimens which grew several bacterial species. We suggest that such disagreement was most likely for urine containing around 10(5) colony-forming units per ml (the culture threshold of positivity) and that these specimens were ones contaminated by urethral or vaginal flora. We propose standard procedures for calibrating and evaluating rapid or automated methods for the detection of significant bacteriuria and have analyzed our results using these procedures. We recommend that identical analyses should be reported for other evaluations of bacteriuria detection methods.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Reference30 articles.

1. Evaluation of an adenosine 5'-triphosphate assay as a screening method to detect significant bacteriuria;Alexander D. N.;J. Clin. Microbiol.,1976

2. Rapid screening for bacteriuria using a particle counter, pulse height analyser, and computer;Alexander M. K.;J. Clin. Pathol.,1981

3. Ansehn S. A. Lundin L. Nilsson and A. Thore. 1979. Detection of bacteriuria by a simplified Iuciferase assay of ATP p. 438-445. In E. Schram and P. Stanley (ed.) Analytical applications of bioluminescence and chemiluminescence. State Printing and Publishing Inc. Westlake Village Calif.

4. Armitage P. 1971. Diagnostic tests p. 433-438. In P. Armitage Statistical methods in medical research. Blackwell Scientific Publications Oxford.

5. .Brown D. F. J. and M. Warner. 1982. Impedance and conductivity methods for detecting bacteriuria p. 171-175. In R. C. Tilton (ed.) Rapid methods and automation in microbiology. American Society for Microbiology Washington D.C.

Cited by 21 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献